=Paper= {{Paper |id=Vol-1337/paper3 |storemode=property |title=Secured SOA for the Safe Interconnection of Medical Devices (Position Paper) |pdfUrl=https://ceur-ws.org/Vol-1337/paper3.pdf |volume=Vol-1337 |dblpUrl=https://dblp.org/rec/conf/se/LeuckerS15 }} ==Secured SOA for the Safe Interconnection of Medical Devices (Position Paper)== https://ceur-ws.org/Vol-1337/paper3.pdf
                   Secured SOA for the Safe Interconnection
                      of Medical Devices (Position Paper)∗

                                   Martin Leucker1 , Malte Schmitz1,2

    1
     Institute for Software Engineering and Programming Languages, Universität zu Lübeck
                            {leucker, mschmitz}@isp.uni-luebeck.de
    2
      Graduate School for Computing in Medicine and Life Sciences, Universität zu Lübeck


            Abstract: This position paper presents a concept on how the dynamic interconnection
            of medical devices in a service oriented architecture (SOA) can be secured based on
            formal interface descriptions. In the framework OpenSDC used in the research project
            OR.NET formal interface descriptions are already used to model medical devices. The
            new approach is to use this formal descriptions to ensure the correct usage of the
            interface at runtime and to support developers in implementing medical devices with
            OpenSDC at the same time.


1 Introduction
As medical devices are safety-critical systems, a risk analysis has to be carried out that
comprehensively investigates the hazards to which a machine can expose the patient dur-
ing surgery. To this end, the interfaces of a machine both to humans but also to other
medical devices have to be precisely defined and measures for any (intentional and unin-
tentional) misuse of the interface have to be taken. In this position paper we suggest the
following approach to ensure the correct employment of a machine’s interface. Starting
from a formal definition of the interface together with constraints about its correct usage,
we suggest to (i) generate a Java interface skeleton from the formal specification to sim-
plify the implementation and thus the programmer’s work, (ii) generate monitors to ensure
the correct usage of the interface and (iii) generate possible mediators whose goal is to
translate between different interface types for which a well-known translation scheme is
available. We are convinced that the presented approach allows the misuse of a given inter-
face to be detected while at the same time offering great flexibility whenever a meaningful
connection of devices is suitable. Using this approach the approval process for the overall
medical device will be simplified significantly, as a prove that identified risk measures are
respected in the implementation is more evident if formal specifications are used to ensure
the correct usage of the interface at runtime.
We outline our ideas within the communication framework OpenSDC [GBF12] used in
the project OR.NET1 as reference implementation of the Open Surgical Communication
Protocol (OSCP). OR.NET addresses the legal and technical challenges of using inter-
        ∗ This
          work is supported in part by the BMBF project OR.NET under number 16KT1231.
1   www.ornet.org




                                                    11
connected medical devices in an operating room (OR) [BB12]. OpenSDC is a Java li-
brary going to be standardized for the communication of interconnected medical devices.
To achieve a flexible, dynamic interconnection of devices a service oriented architecture
(SOA) is used. OpenSDC can be used to implement providers (medical devices) and con-
sumers (medical clients) for this SOA.
An earlier and more abstract version of this concept was given in [KL13] together with an
overview of the legal situation. In this paper we make those abstract ideas more precise.


2 Medical Device/Client Description

                                compatible?
                MC                                              MD

     M           IF                                     IF       M         IF
                         MCDesc           MDDesc
                NW                                     NW                  IO

Figure 1: Simplified layered model of the interconnection. A medical device (MD) is connected
to its peripheral IO, e. g. a keyboard, through an interface (IF). The same way the MD and the
medical client (MC) are connected through an IF with the network stack (NW), but these interfaces
can be generated from the medical device description (MDDesc) or the medical client description
(MCDesc) respectively. The monitors (M) can be generated from the same source.

The medical device information base (MDIB) is an important component of the architec-
ture used in OR.NET. The MDIB is a hierarchical tree-like structure that stores medical
objects. On the lowest layer it describes metrics of a medical device that can be read and
set. The concept of an MDIB is described in ISO/IEEE 11073 [ISO04] as part of the
domain information model for medical devices.
In OR.NET the MDIB consist of two parts: The medical device description (MDDesc)
and the medical device state (MDState). The MDState contains current values of metrics
and the MDDesc provides meta data on these values. These meta data contain information
about the physical quantity and the unit of a metric. Only using both one can interpret the
received values correctly. The units and physical quantities in the MDDesc are described
with code identifiers that are defined in Part 10101: Nomenclature of [ISO04]. In the
OpenSDC implementation the MDDesc can be imported from an XML serialization whose
format is going to be standardized in the OR.NET context.
In the current implementation the client can retrieve the MDDesc from a device and use
this meta data to read and set values in the MDState of the device. In other words the
client can retrieve a formal interface description from the server. To extend this idea we
propose a medical client description (MCDesc) that contains such formal meta data on the
interface of a client. This includes a complete list of metrics set and read by this client and
the physical quantities and units of these metrics.
Given an MDDesc and an MCDesc both serialized as XML one can check the compati-




                                              12
bility of the interfaces of two medical devices. This way an interconnection can only be
allowed if the interfaces of device and client are compatible. At the moment the MDDesc
and MCDesc interface description is stateless, which makes such a check quite trivial, but
limits the expressiveness of these descriptions, too. Consider an instrument that can be in
an ON and an OFF state. While in the ON state the surgeon can apply too much force
and hence the device transitions to the OVERLOAD state. Once in the OVERLOAD state
the surgeon has to turn the device OFF before it is allowed to turn it back ON. To be able
to consider such conditions in the compatibility check, we want to go one step further
and add stateful interface descriptions to the MDDesc and the MCDesc. This could be
achieved using interface automata as proposed in [dAH01]. The authors already present a
ready-to-use compatibility check algorithm.


3 Interface and Monitor Generation
As part of the risk control measures each medical device has to perform a plausibility check
on every incoming parameter [JWHK11]. This holds for input coming from physical
input devices connected directly to the medical device as well as for input coming over
the network. In figure 1 the first case is depicted in the rightmost stack with a keyboard
connected to the IO layer. The plausibility check is implemented manually in the interface
layer and many tests were written to assure the correct implementation of the check.
In the case of interconnected medical devices such a plausibility check is even more impor-
tant, but in the current implementation satisfying its own interface description and check-
ing the correctness of incoming messages is completely left up to the user code. We
therefore propose to automatically generate the interfaces from the already existing formal
interface description in the MDDesc or the new MCDesc. We want to generate some Java
code doing the hard work and some Java interfaces that the user software has to imple-
ment2 . This way technical details like handles of the metrics and most of the complexity
of the highly parallel communication model can be hidden from the user software. This
idea is depicted in the other stacks in figure 1 with the link between the document depicting
the MDDesc and the interface layer.
In addition to the generated Java interface we want to generate monitors from the MDDesc
that assert at runtime that conditions of the interface are satisfied. Of course conditions that
are already enforced by the Java interface are not monitored any more, but a Java inter-
face is not able to express more complex conditions, such as allowed ranges of integer
values. In the case of stateful interface descriptions the generated monitors could check
these conditions, too. Using such monitors provides us with two benefits: We can be sure
that incoming values are always in the range of allowed values and one can trust the com-
patibility check described above. The result of the compatibility check is useless without
the guarantee that every communication partner follows its formal interface description at
runtime. Using the generated monitors on all medical devices in the interconnection, we
can use the compatibility check as a strong argument in risk control.
2 As all parts of OpenSDC are implemented in it, Java is the natural choice when implementing communication
based on OpenSDC.




                                                   13
4 Mediator Synthesis
Last but not least the MDDesc/MCDesc pair can be used to make an interconnection pos-
sible in some cases even if the compatibility check fails. Assume we stored some relations
on physical quantities and units in an ontology. This knowledge can be used to synthesize
a mediator if an MCDesc expects other conditions than the MDDesc provides. For a tech-
nique for automated synthesis of mediators compare [BCIJ13]. Such a mediator would
be a pure software device consisting of a virtual client and a virtual server. The virtual
server provides the device that the MCDesc requests and translates every request to fulfill
the requirements of the MDDesc. A very simple example would be a mediator translat-
ing Celsius to Fahrenheit. More complex examples could involve stateful transitions like
summing up values over time or keeping track of the last values to compute deltas.


5 Summary
With the XML serialization of the MDDesc a formal interface description of medical de-
vices exists, is used in the research project OR.NET and is going to be standardized. Using
and extending this formal interface description we can generate Java interface skeletons to
hide the complexity of OpenSDC from developers of medical devices and generate moni-
tors ensuring the correct usage of the interface at the same time. Using such a technology
the development of and risk analysis for an interconnected medical device will be much
easier.
The next steps include implementing a tool to simplify the modeling of the MDDesc as
well as implementing the generation of the Java skeleton and the monitor from the MD-
Desc. To implement mediator synthesis and the proposed compatibility check we need to
define the proposed MCDesc in more detail.


References
[BB12]   M. Birkle and B. Bergh. OR.NET: Ein Projekt auf dem Weg zur sicheren dynamischen
         Vernetzung in OP und Klinik. In Informatik 2012, volume 208, pages 1235–1236. GI,
         2012.
[BCIJ13] A. Bennaceur, C. Chilton, A. Isberner, and B. Jonsson. Automated Mediator Synthesis:
         Combining Behavioural and Ontological Reasoning. In SEFM, volume 8137 of LNCS,
         pages 274–288. Springer, 2013.
[dAH01] L. de Alfaro and T. Henzinger. Interface automata. In SIGSOFT, pages 109–120. ACM,
         2001.
[GBF12] D. Gregorczyk, T. Bußhaus, and S. Fischer. A Proof of Concept for Medical Device
         Integration using Web Services. In SDD, pages 1–6. IEEE, 2012.
[ISO04]  ISO/IEEE. ISO/IEEE 11073: Health informatics – Point-of-care medical device com-
         munication, 2004.
[JWHK11] C. Johner, S. Wittorf, and M. Hölzer-Klüpfel. Basiswissen Medizinische Software.
         dpunkt.verlag, 2011.