<!DOCTYPE article PUBLIC "-//NLM//DTD JATS (Z39.96) Journal Archiving and Interchange DTD v1.0 20120330//EN" "JATS-archivearticle1.dtd">
<article xmlns:xlink="http://www.w3.org/1999/xlink">
  <front>
    <journal-meta />
    <article-meta>
      <title-group>
        <article-title>Multi-Level Safety Concept for Robot Assisted ENT Surgery für die 11. CURAC Jahrestagung 2012</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>T.M. Williamson</string-name>
          <email>tom.williamson@artorg.unibe.ch</email>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>B. Bell</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>N. Gerber</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>K. Gavaghan</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>J.Anso</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>L. Salas</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>M.Caversaccio</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>S. Weber</string-name>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <aff id="aff0">
          <label>0</label>
          <institution>ARTORG Institute for Biomedical Engineering Research, University of Bern</institution>
          ,
          <country country="CH">Switzerland</country>
        </aff>
        <aff id="aff1">
          <label>1</label>
          <institution>Department ENT Surgery, University Hospital Bern</institution>
          ,
          <country country="CH">Switzerland</country>
        </aff>
        <aff id="aff2">
          <label>2</label>
          <institution>Schlüsselworte: Roboter</institution>
          ,
          <addr-line>Chiurgie, Ohr, Sicherheit</addr-line>
        </aff>
      </contrib-group>
      <fpage>207</fpage>
      <lpage>210</lpage>
      <abstract>
        <p>During image guided interventions the surgeon depends on feedback provided by the surgical support system, in some cases exclusively. Because image guided surgery consists of a number stages during which errors may be introduced, the final system accuracy and robustness can be compromised without knowledge of the user. These errors become particularly important in cases in which extremely high accuracy is required, such as during microsurgery on the lateral skull base. We believe that a navigation system, in addition to providing high accuracy image guidance, could also benefit from additional redundant systems which provide further information about the location of the surgical tool in relation to vital structures. Presented here is a system for robotic cochlear implant surgery which utilizes redundancy and various levels of automation for safety in registration, tracking and tool positioning.</p>
      </abstract>
    </article-meta>
  </front>
  <body>
    <sec id="sec-1">
      <title>Problem</title>
    </sec>
    <sec id="sec-2">
      <title>Methods</title>
      <p>Registration: Patient to pre-operative image registration is one of the major error sources in traditional image guided
surgery [6]. In an effort to reduce the error associated with this process, the robotic system utilizes a semi-automatic,
force based, screw localization technique for registration. First, the user places the robot end effector above each of the
pre-operatively implanted fiducial registration screws using simple force-admittance control. At this point the system
takes over control of the registration process wherein a 1N bias force is applied along the axis of the registration tool,
causing the tool to contact the conical inner surface of the registration screw. Lateral forces are simultaneously
minimized causing the tool to settle in the center of the screw. The position of each screw is then digitized in two separate
coordinate systems: based on the robot encoder values and an optical-tracking system (CamBar B1, Axios3D,
Germany). Screw localization in the image is also completed semi-automatically within a custom-built planning software
(OtoPlan) using a method similar to that described in [7].</p>
      <p>The accuracy of the proposed scheme was validated on a technical phantom: screws were implanted and ground truth
positions measured using a coordinate measurement machine (CMM, Tesa 3D Micro-MS 454, Switzerland). Fiducial
localization error (FLE) was determined by repeatedly registering a single screw from different angles and positions and
measuring the final position of the tool within the screw with the CMM. Target registration error (TRE) was calculated
by registering the technical phantom and measuring designated target screws. The TRE was then defined as the distance
between the (transformed) measured targets and the ground truth CMM target positions.</p>
      <p>Detection of Navigation Errors: Additional error sources such as tool calibration, tracking or process errors (tool
bending, backlash) will also contribute to the ability of the system to accurately reach the defined target position. As
such, the system integrates additional sensor sources for the detection of an incorrect drilling path or potential harm to
anatomy, even in the presence of these errors. These methods include integrated facial nerve monitoring and
forcedensity based pose estimation, as described below.</p>
      <p>Integrated Facial Nerve Monitoring: The facial nerve is the major structure at risk during cochlear implantation
surgery. It is responsible for the innervation of the ipsilateral facial muscles, and damage may result in paralysis. Facial
nerve monitoring systems are common within the ENT operating theater, and many surgeons are familiar with their use.
A Medtronic StimBur drill (StimBur, Medtronic, USA) is directly integrated into the robotic hardware and navigation
software.</p>
      <p>Fig. 1: Tool pose estimation based on the correlation of observed axial drilling force and bone density. Imaging
data is obtained and a region of interest defined (a). A number of candidate trajectories (b) are selected within
this region of interest. Imaging data along these candidate paths is extracted (c), the density along each path
calculated (d) and then correlated with the observed force (e) to provide the estimated position of the tool (f).
Force-Density Pose Estimation: We have recently developed a novel method for the localization of a surgical drill
based on the forces developed during the drilling process. The algorithm correlates these forces with the particular bone
density along a particular trajectory describing a possible drill path. Because the mastoid contains many air cells, any
particular path through the bone exhibits a unique density profile. Thus, the specific force pattern obtained during the
drilling process can be correlated with each candidate density profile obtained preoperatively, and that density profile
with the highest correlation represents the highest probability of representing the actual location of the drill. A diagram
showing the main process steps is shown in Fig. 1.</p>
      <p>Validation of the algorithm was completed by drilling a total of 10 trajectories in 3 human cranial cadaver specimens. A
rigid titanium wire was inserted into each of the trajectories and post-operative imaging completed; the actual final
drilled trajectory as determined from this post-op imaging was then compared to the planned and estimated trajectories.</p>
      <p>208
Control: As a final safety measure, the surgeon remains in control of the robotic system at all times. A 3-state dead man
switch, consisting of wait, proceed and emergency halt conditions, ensure that the robot will not carry out the planned
trajectory unless the surgeon is completely satisfied with the state of the procedure and allows the reversion to a standard
mastoidectomy at any time.</p>
      <p>e)
a)
b)
d)
Registration: The semi-automatic registration technique was found to be highly accurate, with excellent repeatability.
The technique is independent of user variability and can be quickly and reliably accomplished. The RMS FLE of the
scheme was determined to be 0.05 mm, while a mean TRE of 0.101 mm was found.</p>
      <p>Integrated Facial Nerve Monitoring: Facial nerve monitoring was successfully integrated into the existing robotic
system; the integration of facial nerve monitoring together with standard image guidance, into the ENT micro-surgical
system shown in Fig. 2.</p>
      <p>Force-Density Pose Estimation: The force-density estimation algorithm demonstrated a mean tool localization
accuracy of 0.29 mm when utilizing data from the surface of the mastoid to the middle ear canal, 0.38 mm at the level of the
facial nerve and 0.49 mm when utilizing only data from the initial 50% of the drilled path. The results are shown in
Table 1, errors are calculated from the esimated tool pose to the actual position of the tool, as determined from
postoperative imaging.
In this work we have presented a multi-level safety concept for robot assisted ENT surgery and demonstrated the
implementation of this concept within our micro-surgical robotic system. High accuracy semi-automatic registration,
integrated facial nerve monitoring, force-density based pose estimation and high level surgeon control were all integrated
successfully into the hardware and software of the existing robotic system. The system is able to provide a level of
safety higher than that of traditional image guidance through the incorporation of information from a variety of sensor
sources and types. The registration scheme decreases the likelihood of errors during the referencing phase of the
procedure by removing inter-user variability in fiducial screw localization, as well as being highly accurate and repeatable.
Integrated facial nerve monitoring provides the surgeon with feedback similar to that currently utilized within the ENT
operating theater directly in the robot user interface and allows the surgeon to make decisions about the progress of the
procedure based on individual experience. A live sheep study is currently underway to determine more precise
relationships between facial nerve stimulation and probe position. Force based pose estimation provides a means of localizing
the tool within the mastoid even in the presence of registration, calibration, navigation or other errors and can be
completed quickly and accurately. Of particular interest is the accuracy of 0.39 mm at the level of the facial nerve, which can
be utilized to determine if the facial nerve or other nearby anatomy is at risk.</p>
      <p>Finally, as the robotic system does not move without surgeon input, the surgeon remains in control of the system at all
times and can make the decision, based on the variety of feedback provided, whether to continue with the robotic
procedure or revert to a traditional mastoidectomy.</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <ref id="ref1">
        <mixed-citation>
          <string-name>
            <given-names>S.</given-names>
            <surname>Baron</surname>
          </string-name>
          ,
          <string-name>
            <given-names>H.</given-names>
            <surname>Eilers</surname>
          </string-name>
          ,
          <string-name>
            <given-names>B.</given-names>
            <surname>Munske</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J. L.</given-names>
            <surname>Toennies</surname>
          </string-name>
          ,
          <string-name>
            <given-names>R.</given-names>
            <surname>Balachandran</surname>
          </string-name>
          ,
          <string-name>
            <given-names>R. F.</given-names>
            <surname>Labadie</surname>
          </string-name>
          and
          <string-name>
            <given-names>T.</given-names>
            <surname>Ortmaier</surname>
          </string-name>
          ,
          <article-title>"Percutaneous inner-ear access via an image-guide industrial robot system</article-title>
          .,
          <source>" Proceedings of Institution of Mechanical Engineers</source>
          ,
          <string-name>
            <surname>Part</surname>
            <given-names>H</given-names>
          </string-name>
          : Journal of Engineering in Medicine., vol.
          <volume>224</volume>
          , no.
          <issue>5</issue>
          , pp.
          <fpage>633</fpage>
          -
          <lpage>649</lpage>
          ,
          <year>2010</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref2">
        <mixed-citation>
          <string-name>
            <given-names>R.</given-names>
            <surname>Labadie</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J.</given-names>
            <surname>Noble</surname>
          </string-name>
          ,
          <string-name>
            <given-names>B.</given-names>
            <surname>Dawant</surname>
          </string-name>
          ,
          <string-name>
            <given-names>R.</given-names>
            <surname>Balachandran</surname>
          </string-name>
          ,
          <string-name>
            <given-names>O.</given-names>
            <surname>Majdani</surname>
          </string-name>
          and
          <string-name>
            <given-names>J.</given-names>
            <surname>Fitzpatrick</surname>
          </string-name>
          ,
          <article-title>"Clinical validation of percutaneous cochlear implant surgery: initial report</article-title>
          .
          <source>," The Laryngoscope</source>
          , vol.
          <volume>118</volume>
          , no.
          <issue>6</issue>
          , pp.
          <fpage>1031</fpage>
          -
          <lpage>1039</lpage>
          ,
          <year>2008</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref3">
        <mixed-citation>
          <string-name>
            <given-names>T.</given-names>
            <surname>Klenzner</surname>
          </string-name>
          ,
          <string-name>
            <given-names>C.</given-names>
            <surname>Ngan</surname>
          </string-name>
          ,
          <string-name>
            <given-names>F.</given-names>
            <surname>Knapp</surname>
          </string-name>
          ,
          <string-name>
            <given-names>H.</given-names>
            <surname>Knoop</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J.</given-names>
            <surname>Kromeier</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A.</given-names>
            <surname>Aschendorff</surname>
          </string-name>
          ,
          <string-name>
            <given-names>E.</given-names>
            <surname>Papastathopoulos</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J.</given-names>
            <surname>Raczkowsky</surname>
          </string-name>
          ,
          <string-name>
            <given-names>H.</given-names>
            <surname>Wörn</surname>
          </string-name>
          and
          <string-name>
            <given-names>J.</given-names>
            <surname>Schipper</surname>
          </string-name>
          ,
          <article-title>"New strategies for high precision surgery of the temporal bone using a robotic approach for cochlear implantation," European Archives of Oto-Rhino-</article-title>
          <string-name>
            <surname>Laryngology</surname>
          </string-name>
          , vol.
          <volume>266</volume>
          , pp.
          <fpage>955</fpage>
          -
          <lpage>960</lpage>
          ,
          <year>2009</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref4">
        <mixed-citation>
          <string-name>
            <surname>Leinung</surname>
          </string-name>
          ,
          <article-title>"A robot-guided minimally invasive approach for cochlear implant surgery: preliminary results of a temporal bone study," International journal of computer assisted radiology and surgery</article-title>
          , vol.
          <volume>4</volume>
          , pp.
          <fpage>475</fpage>
          -
          <lpage>486</lpage>
          ,
          <year>2009</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref5">
        <mixed-citation>
          <string-name>
            <given-names>B.</given-names>
            <surname>Bell</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Gerber</surname>
          </string-name>
          ,
          <string-name>
            <given-names>K.</given-names>
            <surname>Gavaghan</surname>
          </string-name>
          ,
          <string-name>
            <given-names>C.</given-names>
            <surname>Stieger</surname>
          </string-name>
          ,
          <string-name>
            <given-names>M.</given-names>
            <surname>Caversaccio</surname>
          </string-name>
          and
          <string-name>
            <given-names>S.</given-names>
            <surname>Weber</surname>
          </string-name>
          ,
          <article-title>"Computerassistierte Präzisionschirurgie am Ohr,"</article-title>
          <source>Automatisierungstechik</source>
          , p. In Press,
          <year>2011</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref6">
        <mixed-citation>
          <string-name>
            <given-names>R. F.</given-names>
            <surname>Labadie</surname>
          </string-name>
          ,
          <string-name>
            <given-names>B. M.</given-names>
            <surname>Davis</surname>
          </string-name>
          &amp;
          <string-name>
            <surname>J. M.</surname>
          </string-name>
          <article-title>Fitzpatrick “Image-guided surgery: what is the accuracy?” Current opinion in otolaryngology &amp; head and neck surgery</article-title>
          , vol.
          <volume>13</volume>
          , pp.
          <fpage>27</fpage>
          -
          <lpage>31</lpage>
          , 2005
          <string-name>
            <given-names>G.</given-names>
            <surname>Zheng</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Gerber</surname>
          </string-name>
          ,
          <string-name>
            <given-names>D.</given-names>
            <surname>Widmer</surname>
          </string-name>
          ,
          <string-name>
            <given-names>C.</given-names>
            <surname>Stieger</surname>
          </string-name>
          ,
          <string-name>
            <given-names>M.</given-names>
            <surname>Caversaccio</surname>
          </string-name>
          ,
          <string-name>
            <given-names>L.-P.</given-names>
            <surname>Nolte</surname>
          </string-name>
          ,
          <string-name>
            <given-names>S.</given-names>
            <surname>Weber</surname>
          </string-name>
          ,
          <article-title>"Automated detection of fiducial screws from CT/DVT volume data for image-guided ENT surgery,"</article-title>
          <source>Engineering in Medicine and Biology Society (EMBC)</source>
          ,
          <source>2010 Annual International Conference of the IEEE</source>
          , pp.
          <fpage>2325</fpage>
          -
          <lpage>2328</lpage>
          ,
          <year>2010</year>
          .
        </mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>