        <article-title>Blood Flow Estimation by Means of Intraoperative Rotational Angiographic System</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>Clemens M. Hentschke</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Steffen Serowy</string-name>
          <email>steffen.serowy@ovgu.de</email>
          <xref ref-type="aff" rid="aff1">1</xref>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Georg Rose</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Klaus D. Tönnies</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <aff id="aff0">
          <label>0</label>
          <institution>Department of Simulation and Graphics, University of Magdeburg</institution>
          ,
          <country country="DE">Germany</country>
        </aff>
        <aff id="aff1">
          <label>1</label>
          <institution>Institute for Electronics</institution>
          ,
          <addr-line>Signal Processing, and Communications</addr-line>
          ,
          <institution>University of Magdeburg</institution>
          ,
          <country country="DE">Germany</country>
        </aff>
        <aff id="aff2">
          <label>2</label>
          <institution>Institute of Neuroradiology, University of Magdeburg</institution>
          ,
          <country country="DE">Germany</country>
        </aff>
      </contrib-group>
      <fpage>67</fpage>
      <lpage>70</lpage>
      <abstract>
        <p>We present an algorithm for estimating the blood flow in angiographic image data. In corresponding projective digital subtraction X-ray angiography (2D-DSA) data sets we analyze the correlation between Concentration Time Curves (CTC) of pixels along the vessel centerline. The appropriate spatial information is recovered by applying a 2D-3D registration re-projecting the centerline pixels to the reconstructed 3D X-ray rotation angiography (3D-RA) data of the same object. Ambiguities caused by occluding vessels are resolved by a graph-based approach. Finally, we end up in a framework for the estimation of a real spatial blood flow. This measure is used as boundary condition for blood flow CFD simulations. The algorithm has been tested on phantom data. First plausibility tests of this re-projection method on patient data indicate its ability to also properly function on these data.</p>
      </abstract>
      <kwd-group>
        <kwd>Blood Flow</kwd>
        <kwd>X-ray</kwd>
        <kwd>Angiography</kwd>
        <kwd>Registration</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec id="sec-1">
      <title>Problem</title>
      <sec id="sec-1-1">
        <title>2.1 Measuring projective blood flow</title>
        <p>We aim to determine the mean 2D pixel blood flow velocity from time-dependent 2D-DSA images. These images can
be acquired from arbitrary plane positions by means of modern rotational angiography devices with a sufficiently frame
rate of at least 30 frames per second (fps). Blood flow is measured by detection of contrast agent (CA) and its
propagation through time. Since injection of the CA strongly disturbs the blood flow propagation, the estimation of the true
blood flow velocity by bolus tracking requires a linear correction factor [1]
A 2D-MAX image is produced by computing a maximum intensity projection along the time axis to ensure that vessel
pixels are clearly distinguishable from background pixels. A ROI is chosen manually, which contains the feeding vessel
to an aneurysm. The flow in the vessel is characterized by measuring the blood flow along its centerline. The centerline
is extracted by threshold segmentation of the 2D-MAX image followed by a thinning step that produces a centerline
with a width of 1 pixel and 8-cell neighborhood. The centerline extraction is semi-automatic and requires little
userinteraction. Distance information along a centerline is generated by tracking a path from the starting point to the
endpoint.</p>
        <p>Once the vessel and its centerline are segmented, projected blood flow velocity can be estimated. There are two major
approaches to compute blood flow propagation from angiographic images [1]. Concentration Time curves (CTC) track
the bolus propagation by observing the change of pixel intensity while Concentration Distance Curves (CDC) track the
front propagation of the bolus. We have chosen the first approach, because we are interested in a mean velocity along
the vessel centerline. Due to pulsative behavior of the blood flow, this can be measured integrally more precisely by
CTC. Furthermore, CTC take all time frames at all locations into account, while CDC relies on time frames where the
bolus propagation is visible.</p>
        <p>Since blood flow is assumed to have constant velocity, it can be derived from observing the displacement of the
intensity profile of contrast agent along the vessel centerline. Propagation of flow at a position pi with respect to a position p0
on the vessel centerline is measured by the delay δt of the two corresponding CTC and the known distance Δs between
p0 and pi. The velocity is then v = Δs/ δt. The delay is given by the maximum position of the cross correlation of the
CTC at position p0 and the CTC at position pi. This is measured for all possible pairs of positions.
As the CTC only provides discrete intensity information at each time an image has been acquired, the cross-correlation
function is a discrete function as well. For a more accurate determination of the maximum position, we use a 4th order
polynomial interpolation scheme around the discrete maximum position to estimate the maximum position in
continuous space. This step is important in case of a short centerline, where the expected values of the position of the
maximum are small.</p>
        <p>The continuous maximum positions for each centerline pixel provide information about δt and Δs. In order to find a
mean blood flow velocity, we perform a linear interpolation of all values. The slope of the interpolation function yields
the reciprocal pixel velocity. As indicated before, a linear correction factor c has to be included to estimate the correct
blood flow velocity from measured bolus propagation. We estimate the factor experimentally by injecting different
volume amounts into the cerebral phantom (Section 3). The undisturbed blood flow is computed by taking advantage of
the linear relationship between contrast agent quantity and measured blood flow velocity. The measured blood flow
velocity is plotted against the quantity of the injected CA. A linear regression line extrapolates the true value of blood
flow that is the value for a quantity of 0 ml/s. The phantom correction factor estimated for the phantom is then
transferred to the patient data. As the amount of CA is known, the correction factor can be estimated. However, the
correction factor is also dependent on the vessel diameter [1]. We assume that the diameter of the feeding vessels does only
vary little. This is the case for vessels in the Circle of Willis, which we in this paper focus on..</p>
      </sec>
      <sec id="sec-1-2">
        <title>2.2 Recover spatial information from projective images</title>
        <p>Projective blood flow is measured based on a 2D-DSA image sequence at vessel centerline pixels. For recovering
metric information we need to assign 3D world coordinates to every pixel of the 2D centerline. Therefore, a re-projection
of 2D-DSA data into 3D-RA volume is required. Spatial alignment of the two data sets is given, since a prior 2D-3D
registration presented in [5] provides the transformation parameters.</p>
        <p>We employ a ray-shooting approach to backproject the projective information from 2D-DSA to registered 3D vessels
extracted from 3D-RA. Ambiguous mappings occur if a ray intersects more than one vessel in the 3D scene. We solve
the mapping problem by transforming it into a graph-based problem. An undirected graph is built to describe all
possible 3D correspondences of the 2D centerline pixels. All vertices that share a common parent 2D centerline neighbor
pixel are connected. Costs, associated to the edges of the graph, are defined by intensity, position and connectivity. An
optimal path is computed by applying Dijkstra's algorithm.
Finally, we are able to determine the length of the 3D centerline by measuring the Euclidian distance between the
centerline nodes. Hence, velocities in [px/s] as a result of the computation in the 2D-DSA data set can be transferred into
flow in [mm/s] or [ml/s], respectively, by incorporating a segmented 3D-RA image.</p>
      </sec>
    </sec>
    <sec id="sec-2">
      <title>Results</title>
      <p>We evaluated our method on phantom as well as on patient data. Two different information provided ground truth to
validate our measurements on the phantom: Laser Doppler Velocimetry (LDV) and the total delivery rate of the used
pump. In the case of patient data, only plausibility tests could be performed as ground truth information was not
available. However, the main focus of this paper is to prove the general functionality of the presented framework.
We used a cerebral vessel silicone phantom that includes three aneurysms (Elastrat H+N-R-A-EV-003). The fluid was
pulsatively pumped through the artificial vessels by means of a pump. For the experimental setup and further
information about LDV, we refer to [6].</p>
      <p>LDV measurements led to a mean flow velocity of 226 mm/s, measured in the investigated. The known delivery rate of
the pump was 5.25 ml/s and the mean diameter of the feeding vessel was 5.3 mm. It results in a mean volume flow
within the vessel of 229 mm/s. Hence, due to the uncertainties of the given parameters the ground truth differ by
approximately 1.3 %.</p>
      <p>We measured the mean projective blood flow velocity with 2D-DSA images acquired from 2 different angles. By our
method, we measured a blood flow of 1075 px/s for angle 1 and 1092 px/s for angle 2. The centerline had a projective
length of 434.8 px and 526.6 px, respectively. The spatial length was 107.6 mm and 98.4 mm. Division of the spatial
length by the pixel length yields a total pixel spacing. Multiplying this value by the projective pixel flow, it results in a
volume flow of 266 mm/s for the projective data set for angle 1 and 204 mm/s for the projective data set for angle 2.
Hence, the deviation was 17.7 % against the LDV measurements and 16.2 % compared with the the pump delivery rate
for the first data set. The precision of the flow results for the second data set was 9.7 % and 10.9 %, respectively. For a
patient data set, we determined a blood volume flow of 2.21 ml/s in a cerebral vessel with a diameter of 4.6 mm.
As our algorithm is designed for the usage of real patient data, we also performed experiments on five patient data sets.
Due to the lack of ground truth, we carried out plausibility tests of our spatial recovery approach (section 2.2). In two
bi-plane 2D-DSA data sets (difference of the plane angles is approximately 90°), the projection of a 3D centerline has a
different path and length. We used this fact to re-project the manually chosen 2D centerlines that share a common 3D
centerline. Hence, the re-projected length of the centerline is ideally the same for both 2D-DSA data sets. We have
manually chosen feeding vessels that are defined by a start and end point. Both points are defined by salient images
features like bifurcations and aneurysms. Usually, feeding vessels are larger vessels as the probability of aneurysms
formation is higher. As measure, we used the deviation on a percentage basis from the length of the spatial recovered 3D
centerlines. We measured a deviation between 0.3 % and 6.9 % with a mean deviation of 3.9 %. This is accurate under our
assumptions as the 2D centerlines differ quite strong in terms of path length. Additionally, the re-projection is disturbed
by close-by vessels and vessels occluding the vessel of interest.
In order to support physicians in the treatment of intracranial aneurysms the qualitative information on blood flow in
the feeding vessels is of high importance.</p>
      <p>We presented a method to measure the blood flow velocity in angiographic image data. Projective pixel blood flow
velocity was extracted from 2D-DSA by measuring the blood flow at vessel centerline positions. Spatial information was
included by incorporating 3D-RA data. A 2D-3D registration and a consecutive re-projection provide the
transformation of pixel data to 3D coordinates. A graph-based scheme was introduced to resolve ambiguities caused by occluding
vessels.</p>
      <p>We tested our algorithm on a cerebral vessel phantom and proved the quality and the functionality of our approach.
With respect to our assumptions, we are able to compute the blood flow velocity with good accuracy. Compared to
literature, we presented several improvements in different scopes. The projective blood flow propagation is computed by
comparing Concentration Time Curves of all vessel centerline pixels rather than taking just a subset of pixels into
account. To recover spatial information from projective images, we use a re-projection approach that solves projection
ambiguities by estimating an optimal cost based path through a graph that is built by registration knowledge. This is a
rather straightforward, but easily extendible method to map positions from a 2D-DSA image to a 3D-RA volume.
The results lead to an improvement in CFD simulations as the inflow blood velocity is not estimated from patient data,
but usually given by a uniform distribution according to literature. With our approach, for each phantom or patient a
specific boundary condition for the CFD simulation can be applied. This is especially interesting in the case of patient
data. However, the evaluation of our results and comparison with other methods is difficult since accurate quantitative
measurements are typically not given or, if available, not directly comparable as the experiments and ground truth data
differ. As an example, the reported accuracy in [7] is similar to our precision. With respect to our measurements, we
expect that a better utilization of the relationship of bi-plane projection images into the process will lead to a more precise
estimation of the blood flow velocity.</p>
      <p>For the future, we plan to test our algorithm with further patient data. In this context, the transfer of the correction
factor to recover true blood flow from measured bolus propagation is subject to further investigation. Additionally, we plan
to test the influence of parameters like position and length of the chosen vessel centerline on our algorithm.
5</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <ref id="ref1">
        <mixed-citation>
          <string-name>
            <surname>Shpifoygel</surname>
            <given-names>SD</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Close</surname>
            <given-names>RA</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Valentino</surname>
            <given-names>DJ</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Duckwiler</surname>
            <given-names>GR</given-names>
          </string-name>
          ,
          <article-title>X-ray videodensitometric methods for blood flow and velocity measurement: A critical review of literature</article-title>
          .
          <source>Medical Physics</source>
          ,
          <year>2000</year>
          ;
          <volume>27</volume>
          :
          <string-name>
            <surname>2008 Schmitt</surname>
            <given-names>H</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Grass</surname>
            <given-names>M</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Suurmond</surname>
            <given-names>R</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Köhler</surname>
            <given-names>T</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Rasche</surname>
            <given-names>V</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Hähnel</surname>
            <given-names>S</given-names>
          </string-name>
          , et al.
          <article-title>Reconstruction of blood propagation in Three-dimensional rotational X-ray angiography (3D-RA)</article-title>
          .
          <source>Computerized Medical Imaging and Graphics</source>
          .
          <year>2005</year>
          ;
          <volume>29</volume>
          (
          <issue>7</issue>
          ):
          <fpage>507</fpage>
          -
          <lpage>520</lpage>
          .
        </mixed-citation>
      </ref>
      <ref id="ref2">
        <mixed-citation>
          <string-name>
            <surname>Liu</surname>
            <given-names>A</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Bullit</surname>
            <given-names>E</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Pizer</surname>
            <given-names>SM</given-names>
          </string-name>
          ,
          <article-title>3D/2D registration via skeletal near projective invariance in tubular objects</article-title>
          .
          <source>Lecture Notes in Computer Science</source>
          .
          <year>1998</year>
          ; p.
          <fpage>952</fpage>
          -963
          <string-name>
            <surname>Rholfing</surname>
            <given-names>T</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Denzler</surname>
            <given-names>J</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Grassl</surname>
            <given-names>C</given-names>
          </string-name>
          , et al.
          <article-title>Markerless real-time 3-D target region tracking by motion backprojection from projection images</article-title>
          .
          <source>IEEE transactions on medical imaging</source>
          .
          <year>2005</year>
          ;
          <volume>24</volume>
          (
          <issue>11</issue>
          ):1455
          <string-name>
            <surname>Hentschke</surname>
            <given-names>CM</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Tönnies</surname>
            <given-names>KD</given-names>
          </string-name>
          ,
          <article-title>Automatic 2D-3D Registration of Cerebral DSA Data Sets</article-title>
          .
          <source>In: Proc. BVM</source>
          <year>2010</year>
          .
        </mixed-citation>
      </ref>
      <ref id="ref3">
        <mixed-citation>
          <string-name>
            <surname>Aachen; 2010 Bölke</surname>
            <given-names>T</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Seshandhri</surname>
            <given-names>S</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Gürvit</surname>
            <given-names>O</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Bade</surname>
            <given-names>R</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Preim</surname>
            <given-names>B</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Janiga</surname>
            <given-names>G</given-names>
          </string-name>
          , et al.
          <article-title>Phantom based flow analysis by means of dynamic angiography, CFD and laser-doppler-velocimetry</article-title>
          .
          <source>In: IEEE Nuclear Science Symposium Conference Record</source>
          ,
          <year>2007</year>
          , NSS'
          <fpage>07</fpage>
          , Vol
          <volume>5</volume>
          ;
          <year>2007</year>
          , p.
          <fpage>3440</fpage>
          -3445
          <string-name>
            <surname>Waechter</surname>
            <given-names>I</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Bredno</surname>
            <given-names>J</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Barratt</surname>
            <given-names>DC</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Weese</surname>
            <given-names>J</given-names>
          </string-name>
          ,
          <string-name>
            <surname>Hawkes</surname>
            <given-names>DJ</given-names>
          </string-name>
          .
          <article-title>Quantification of blood flow from rotational angiography</article-title>
          .
        </mixed-citation>
      </ref>
      <ref id="ref4">
        <mixed-citation>
          <source>Lecture Notes in Computer Science</source>
          .
          <year>2007</year>
          ;
          <volume>4791</volume>
          :
          <fpage>634</fpage>
          .
        </mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>