        <article-title>An Accuracy Approach to Robotic Microsurgery in the Ear</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>B. Bell</string-name>
          <email>brett.bell@istb.unibe.ch</email>
          <xref ref-type="aff" rid="aff2">2</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>J.Salzmann</string-name>
        </contrib>
        <contrib contrib-type="author">
          <string-name>E.Nielsen</string-name>
          <xref ref-type="aff" rid="aff3">3</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>N.Gerber</string-name>
        </contrib>
        <contrib contrib-type="author">
          <string-name>G.Zheng</string-name>
        </contrib>
        <contrib contrib-type="author">
          <string-name>L.Nolte</string-name>
        </contrib>
        <contrib contrib-type="author">
          <string-name>C.Stieger</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>M.Caversaccio</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>S. Weber</string-name>
        </contrib>
        <aff id="aff0">
          <label>0</label>
          <institution>ARTORG, University of Bern</institution>
          ,
          <country country="CH">Switzerland</country>
        </aff>
        <aff id="aff1">
          <label>1</label>
          <institution>Department ENT Surgery, University Hospital Bern</institution>
          ,
          <country country="CH">Switzerland</country>
        </aff>
        <aff id="aff2">
          <label>2</label>
          <institution>Institute for Surgical Technologies and Biomechanics, University of Bern</institution>
          ,
          <country country="CH">Switzerland</country>
        </aff>
        <aff id="aff3">
          <label>3</label>
          <institution>Interstate Technical School</institution>
          ,
          <addr-line>Buchs</addr-line>
          <country country="CH">Switzerland</country>
        </aff>
      </contrib-group>
      <fpage>77</fpage>
      <lpage>81</lpage>
      <abstract>
        <p>Concerns of rising healthcare costs and the ever increasing desire to improve surgical outcome have motivated the development of a new robotic assisted surgical procedure for the implantation of artificial hearing devices (AHDs). This paper describes our efforts to enable minimally invasive, cost effective surgery for the implantation of AHDs. We approach this problem with a fundamental goal to reduce errors from every component of the surgical workflow from imaging and trajectory planning to patient tracking and robot development. These efforts were successful in reducing overall system error to a previously unattained level.</p>
      </abstract>
      <kwd-group>
        <kwd>Robot</kwd>
        <kwd>Surgery</kwd>
        <kwd>Ear</kwd>
        <kwd>Instrument Guidance</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec id="sec-1">
      <title>Problem</title>
    </sec>
    <sec id="sec-2">
      <title>Methods</title>
      <p>markers placed prior to imaging in both CT and DVT machines. Low and high resolution scans were obtained from
each machine for a side by side comparison by an oto-neuroradiologist.</p>
      <p>Marker Detection: Typical identification of fiducial marker locations is typically performed by hand-picking marker
center points in image planes. This procedure results in voxel sized accuracy and is highly dependent on user
experience and skill. In order to improve this process, a semi automated algorithm was developed to match 3D CAD data
against image data. Once the two data sets were registered, the centroid of the CAD data is then transformed to the
image coordinate system and used as the fiducial point [4].</p>
      <p>Tracking and Registration: Typical image guides surgeries rely on a tracking device to register the patient to the
image data and ultimately guide the surgeon to the proper tool pose. However, optical tracking systems have a limited
accuracy. For this reason, we use the robotic manipulator to register fiducial marker positions rather than relying on
optical tracking systems. However, optical tracking is retained in the system as a safety device, essentially confirming the
position of the robot in relation to the patient. Thus, we identified a new micro-optical-measurement system (Cambar
B1, Axios GmbH, Germany) with higher spatial resolution. The suitability of this measurement system for our
application was assessed and compared to an NDI Optotrak (NDI, Waterloo, ON, Canada) system as a benchmark. This was
done by moving trackers precisely in 3D space using a mechanical positioning device.</p>
      <p>Robotic Manipulator: A newly designed robotic manipulator was created specifically for the task of ORL
microsurgery. The characteristics of the robot are summarized in table 3. Most notably, much effort was expended in making the
robot as light as possible so that it could be easily mounted directly to the operating table, making OR integration as
simple as possible. This feature also has the added benefit of reducing relative motion between the patient and the
robot. Additionally, the robot incorporates a force-torque sensor mounted between the end effector and the surgical
instrument. The sensor also functions as a safety device by limiting tool tip forces encountered during surgery. Finally,
the sensor provides a method for the surgeon to interact directly with the manipulator and essentially guide the robot by
hand.
Accuracy Assessment: The accuracy of the entire system was assessed by performing the entire clinical workflow on a
technical phantom. The phantom consisted of a polyurethane foam block fitted with fiducial markers (figure 2). This
was scanned using a CT scanner, and the resulting data was processed for multiple drilling operations. Trajectories
originating on the top surface of the phantom and ending at internal fiducial markers were planned and sent to the robot
system. Once this process was completed, the phantom was registered using the robotic manipulator. This registration
process was performed by ‘driving’ the robot tool tip to the fiducial markers located on the phantom, and recording
each position in the robot coordinate frame. Following registration, the robot was commanded to automatically perform
a drilling operation. The registration and drilling process was repeated 7 times. Finally, the phantom was reimaged and
the resulting drill trajectories were segmented and compared to the planned trajectories.
3</p>
    </sec>
    <sec id="sec-3">
      <title>Results</title>
      <p>Imaging: Lack of ground truth makes evaluation of the absolute accuracy of imaging and image registration
procedures difficult. However, quantitatively, DVT images provide much higher spatial resolution, though, qualitatively,
they suffer from lower contrast and increased noise. In spite of these problems, an oto-neuroradiological expert found
the images to be qualitatively sufficient and the benefit of increased resolution overrides the loss of contrast.
Marker Detection: Use of the semi-automatic registration technique was clearly advantageous compared to manual
marker detection. Results of the semi-automatic marker detection procedure are summarized in Figure 1: The robot
system includes: Axios micro-measurement camera (A), robotic manipulator (B), end effector with force-torque
sensor (C), and patient tracking devices (D-micro, E-conventional).</p>
      <p>Table 2, where it is readily apparent that the semi-automatic procedure reduced the standard error from a maximum of
0.59mm for manually selected markers in CT data to a minimum of 0.011 for the semi-automatic detection in DVT data.
This result highlights the synergistic effect of higher resolution datasets and the semiautomatic detection technique.
The newly developed robotic manipulator was successfully used to drill holes in a phantom representing the temporal
bone. Its light weight makes it easy to handle and mount to the OR table, and because all the motor drives are integrated
into the robot, it only requires two cables to communicate with and control (summarized in Table 3). The small
workspace of the camera (Figure 1A) allows it to be mounted directly to the robot base (B), while the small tracking markers
are attached to the patient in a pattern (D). The rigid body depicted by the markers is much smaller than traditional
tracking markers (E).</p>
      <p>Drilling Experiment: Figure 2 shows a segmentation of the technical phantom in the post operative CT scan. Pre and
post operative trajectories were compared to analyze the accuracy of surgical assist system. The drilled segment of the
model had a thickness of 8 mm. These holes were segmented in the model and represented with a blue shade as seen in
figure 2. A line was fit to the center of each segmented portion, and compared to the planned trajectory (white lines).
Using this method, it was possible to calculate the angle and Euclidian distance between the planned and achieved
trajectories at the entry point of the block as reported in table 4. The max/min distances were measured at 0.39/0.05 mm
respectively, whereas max/min angles were0.52/2.17°.
Min
Max
Average
Std dev
Median
1
2
In this work, we have represented a computer controlled surgical manipulator specifically designed to aid surgeons in
implanting AHDs. In contrast to rigid drill guides, this approach benefits from the ability of the surgeon to change the
drill trajectory at any time, even drilling multiple trajectories if the surgeon were to implant multiple electrode arrays.
Furthermore, the future development of a milling mode will allow the creation of complex forms in addition to a simple
drill path.</p>
      <p>One major drawback of the current robotic system is the mounting fixture. Currently, three 7 DoF arms are used to fix
the robot to the OR table which allows the robot to be mounted rigidly (&gt;10 as rigid as the robot) in a more or less
arbitrary position. However, this fixation system is too difficult to mount in a surgical setting. Thus, a simple light weight
mounting structure is being designed.</p>
      <p>By systematically addressing each stage in the surgical procedure (imaging, trajectory planning/registration, tracking,
robotics), we have successfully met the 0.5mm goal to safely access the cochlea (avoiding nerves) through a single drill
pass. Additionally, the demonstrated accuracy exceeds previous attempts by Majdani et al [3]. This achievement is most
likely attributable to using the robot as a registration tool. This technique essentially eliminates the errors due to an
optical tracking system, which are 0.25 mm at a minimum. Furthermore, because registration occurs in the robot
coordinate system, this tends to reduce the effect of small calibration errors which may be present in the robot itself.
In addition to the ability to drill a precisely defined access, this robot system also incorporates several distinct
advantages such as a force-torque sensor which allows the surgeon to interact directly with the robot to control its motion
during drilling and milling procedures. Additionally, this force sensing will allow the robot to automatically detect
tissue boundaries similar to the method used by Brett et al. in cochleostomies [5, 6]. Additionally, the addition of haptic
feedback control is possible which will ultimately enable surgeons to extend current minimally invasive procedures to
previously unexplored opportunities.
5</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <ref id="ref1">
        <mixed-citation>
          <string-name>
            <surname>Schipper</surname>
            ,
            <given-names>J.</given-names>
          </string-name>
          , et al.,
          <article-title>[Navigation-controlled cochleostomy. Is an improvement in the quality of results for cochlear implant surgery possible?]</article-title>
          . HNO,
          <year>2004</year>
          .
          <volume>52</volume>
          (
          <issue>4</issue>
          ): p.
          <fpage>329</fpage>
          -
          <lpage>35</lpage>
          .
        </mixed-citation>
      </ref>
      <ref id="ref2">
        <mixed-citation>
          <string-name>
            <surname>Labadie</surname>
            ,
            <given-names>R.F.</given-names>
          </string-name>
          , et al.,
          <article-title>Minimally invasive, image-guided, facial-recess approach to the middle ear: demonstration of the concept of percutaneous cochlear access in vitro</article-title>
          .
          <source>Otol Neurotol</source>
          ,
          <year>2005</year>
          .
          <volume>26</volume>
          (
          <issue>4</issue>
          ): p.
          <fpage>557</fpage>
          -
          <lpage>62</lpage>
          .
        </mixed-citation>
      </ref>
      <ref id="ref3">
        <mixed-citation>
          <string-name>
            <surname>Majdani</surname>
            ,
            <given-names>O.</given-names>
          </string-name>
          , et al.,
          <article-title>A robot-guided minimally invasive approach for cochlear implant surgery: preliminary results of a temporal bone study</article-title>
          .
          <source>Int J Comput Assist Radiol Surg</source>
          ,
          <year>2009</year>
          .
          <volume>4</volume>
          (
          <issue>5</issue>
          ): p.
          <fpage>475</fpage>
          -
          <lpage>86</lpage>
          .
        </mixed-citation>
      </ref>
      <ref id="ref4">
        <mixed-citation>
          <string-name>
            <surname>Zheng</surname>
            ,
            <given-names>G.</given-names>
          </string-name>
          , et al.,
          <source>Automated Detection of Fiducial Screws</source>
          from CT/
          <article-title>DVT Volume Data for Image-Guided ENT Surgery</article-title>
          , in Engineering in Medicine and
          <source>Biology Society</source>
          .
          <year>2010</year>
          , IEEE: Buenos Aires.
        </mixed-citation>
      </ref>
      <ref id="ref5">
        <mixed-citation>
          <string-name>
            <surname>Brett</surname>
            ,
            <given-names>P.N.</given-names>
          </string-name>
          , et al.,
          <article-title>A surgical robot for cochleostomy</article-title>
          .
          <source>Conf Proc IEEE Eng Med Biol Soc</source>
          ,
          <year>2007</year>
          .
          <year>2007</year>
          : p.
          <fpage>1229</fpage>
          -
          <lpage>32</lpage>
          .
        </mixed-citation>
      </ref>
      <ref id="ref6">
        <mixed-citation>
          <string-name>
            <surname>Coulson</surname>
            ,
            <given-names>C.J.</given-names>
          </string-name>
          , et al.,
          <article-title>An autonomous surgical robot for drilling a cochleostomy: preliminary porcine trial</article-title>
          .
          <source>Clin Otolaryngol</source>
          ,
          <year>2008</year>
          .
          <volume>33</volume>
          (
          <issue>4</issue>
          ): p.
          <fpage>343</fpage>
          -
          <lpage>7</lpage>
          .
        </mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>