=Paper= {{Paper |id=Vol-1477/paper55 |storemode=property |title=A navigated 4-DOF mechanical aiming device for percutaneous ablation of liver tumours |pdfUrl=https://ceur-ws.org/Vol-1477/Proceedings_CURAC_2013_Paper_55.pdf |volume=Vol-1477 |dblpUrl=https://dblp.org/rec/conf/curac/ToporekPLGMDW13 }} ==A navigated 4-DOF mechanical aiming device for percutaneous ablation of liver tumours== https://ceur-ws.org/Vol-1477/Proceedings_CURAC_2013_Paper_55.pdf
A navigated 4-DOF mechanical aiming device for percutaneous ablation
                        of liver tumours

                   G. Toporek¹, M. Peterhans3, Y. Lachenal2, B. Guiu2, A. Monier2, A. Denys2, S. Weber¹

                 ¹ ARTORG Center for Biomedical Engineering Research, University of Bern, Switzerland
          2
              Department of Radiology and Interventional Radiology, Lausanne University Hospital, Switzerland
                                          3
                                            CAScination AG, Bern, Switzerland


                                 Kontakt: grzegorz.toporek@artorg.unibe.ch

Abstract:
During percutaneous interventional procedures, the orientation and position and of a needle-like instrument can be
constrained relative to the patient’s coordinate system using a variety of aiming devices. Physicians align the aiming
device in space according to a pre-defined plan and advance the instrument without excessive bending, resulting in a
higher targeting accuracy. However, available systems have been developed for non-stereotactic procedures and their
use in combination with optical tracking systems is limited due to a tedious alignment process, non-intuitive usage and
operational complexity. Thus, an intuitive 4-DOF aiming device, for specific integration with an optical navigation sys-
tem, was developed and evaluated. Effective accuracy was assessed during 20 image-guided needle punctures (using an
anthropomorphic torso phantom) carried out by three interventional radiologists. Additionally, physicians assessed us-
ability and design properties using a questionnaire. In this work, we report on our findings in terms of the learning
curve for maneuvering the device and adequacy of needle stabilization.

Keywords: interventional radiology, percutaneous liver ablation, navigation system, aiming device

Problem

For patients with early-stage hepatocellular carcinoma, thermal ablation is an established non-surgical treatment method
that provides minimal invasiveness and recovery time when compared to both open surgical and laparoscopic resection
[1]. The clinical outcomes of minimally-invasive thermal ablations are, however, largely dependent on accurate implan-
tation of single or multiple treatment applicators into the tumor [2]. To decrease the dependence of targeting accuracy
on interventionist’s spatial orientation, skills and experience, stereotactic navigation systems have been introduced. Pre-
vious studies have shown that navigated thermal ablations may improve therapeutic outputs and achieve equivalent re-
sults to open liver surgery [3].
Apart from soft tissue deformation, the major challenges during an image-guided procedure are needle stabilization dur-
ing the insertion process and minimization of needle deflection. While electromagnetic tracking systems allow the posi-
tion of the needle tip to be precisely tracked using single coil electromagnetic sensors [4], optical tracking allows only
for the tracking of an instruments distal end. The main shortcoming of electromagnetic navigation is that the overall
tracking, and subsequently targeting accuracy is subjected to field distortions caused by ferromagnetic objects (CT ta-
ble, gantry) and strongly depends on the effective emitter-sensor distance (electromagnetic field strength is inversely
proportional to the cube of the distance). Optical navigation systems, which have larger measurement volumes and a
high tracking precision seem to be more reliable in CT suites [5]. However, optical tracking are not without drawbacks,
including a line-of-sight requirement and the assumption that tracked instruments are rigid bodies. This last assumption
is not correct for the majority of ablation treatment applicators, which may deform while inserting into the soft tissue.
To provide the best possible stiffness while maintaining needle position, several aiming devices have been introduced
[6], [7], [8]. The clinically available SeeStar (Radi, Uppsala, Sweden) is a needle guide which is mounted on the pa-
tient’s skin; no computer guidance is provided in combination with the device, forcing the radiologist to use and itera-
tive adjustment approach based on several control scans. To decrease the number of required CT scans during the inter-
vention, a navigated aiming device has been also developed [7] and compared to navigated free-hand technique on a
non-rigid liver phantom [9]. The lateral components of the errors measured at the needle tip were remarkably lower
when an aiming device was used (2.3±1.3 mm) than in the case of free-hand punctures (4.2±2.0 mm), however use of
this device required both a long learning process and demonstrated long alignment times. To decrease the time taken for
needle alignment, a robotic system was employed, providing remote, semiautomatic needle alignment with a joystick
and thereby preventing radiation burden to medical stuff in fluoroscopy-guided ablations [10]. Although operator errors


                                                           234
and intervention time are reduced, the clinical acceptance of robotic needle guidance is currently low due to the high
complexity and costs of such systems. To decrease this complexity, a novel 4-degree of freedom (DOF) aiming device
equipped with adjustable needle holder (iSYS Medizintechnik GmbH, Austria) was developed. Two large screws placed
above and below aiming device allow an intuitive way to change needle orientation and translation. Despite that, the
usage of such a device requires an iterative needle insertion that must be supervised with several controls CT scans.
Subsequently, the aim of this study was to integrate this 4-DOF aiming device into an optical navigation system dedi-
cated for interventional radiology (IR) as well as qualitatively assessing its usability parameters in a phantom trial using
a questionnaire.

Methods

A stereotactic image-guidance system (CAS-One IR, CAScination AG, Switzerland) dedicated for percutaneous radio-
logical interventions on the liver has been developed and evaluated in several phantom studies [9], [11], [12]. The base
of a navigation system is an optical position measurement system (NDI Vicra, Northern Digital, Canada) that tracks the
needle holder of the aiming device (iSYS Medizintechnik GmbH, Austria) through a dedicated custom-made optical
reference attached with retro-reflective passive spheres (Figure 1, left). The geometry of the marker shield was opti-
mized for continuous line-of-sight during the needle insertion (avoiding marker coverage by the needle) as well as for
visibility under the range of insertion angles required for needle placements on the liver (lateral and posterior trajecto-
ries). The aiming device is pre-calibrated and attached via a mechanical arm (iSYS Medizintechnik GmbH, Austria) to a
custom-made fiberglass plate situated under the patient. Additionally, a set of single retro-reflective marker spheres
(SM) is fixed onto the patient skin around the area of the expected needle incisions to enable real-time patient tracking
and respiratory motion control [13].
In the standard hospital workflow the patient is placed on the CT table with their head inside the gantry, providing ac-
cess for anesthesia from the other side of the gantry. The navigation system is placed opposite the radiologist while the
tracking camera is positioned above the patient, directed caudally (Figure 1, right). The intervention then begins with
the radiologist attaching the single-use needle holder to the aiming device; the aiming device and the mechanical arm
are then draped with a sterile cover. After an initial imaging sequence, the image data is transferred and needle place-
ments are planned on the navigation system. Subsequently, the radiologist places the aiming device in an approximate




          Figure 1: A navigated 4-DOF aiming device with attached marker shield placed above the torso
          phantom (left) that has been tested with navigation system in a realistic radiological setup (right).

position near the expected needle insertion location using guidance information provided by the navigation system. A
final precise alignment of the aiming devices position and orientation is performed under visual control by viewing the
navigation systems targeting viewer. Once the aiming device is precisely aligned, the distance between the reference
mark and the target are displayed on the navigation system. A biocompatible pen is used to put a mark of the required
distance on the needle; the needle is then inserted to this depth.
In this experiment, one series of 20 needle punctures were performed on an anthropomorphic torso phantom, in which
liver tissue was simulated by polymer foam. The end of five 0.018-inch fibered platinum coils (length 20 mm, diameter
3 mm), were randomly inserted in the phantom, and served as targets. A navigation dataset of the phantom was acquired
(128-slice Philips Ingenuity, Voxel size: 0.79 × 0.79 × 0.80 mm3). Each target was reached by a 17 gauge, 168 mm long
co-axial needle (BioPince, Angiotech, Canda) from two different trajectories, varying in length, orientation, position of
incision point (lateral, posterior, anterior) and plane (oblique, in-plane). The trajectories were planned at extreme posi-
tions in order to challenge the usability of the aiming device.


                                                            235
After completion of the experiment, physicians with varying experience levels (experienced: n=2, novice: n=1) were
asked to fill out a questionnaire regarding the benefit and feasibility of using the 4-DOF aiming device with the optical
navigation system. The questionnaire included questions referring to the usability of the aiming device in interventional
radiology (IR) suite, its movement capabilities, the advantages and disadvantages of needle stabilization during inser-
tion, as well as questions concerning mechanical design. The closed-end questions could be answered based on a fol-
lowing scale: “Strongly disagree (0%)”, “Disagree (30%)”, “Agree (70%)”, “Strongly agree (100%)”. Average agree-
ments above or equal 70% were considered as positive indication for given statements.

3        Results

10 needle trajectories (average length 119.0 mm) were planned on the CT images. Five different targets were reached
from different orientations using the navigated 4-DOF aiming device. One of twenty trajectories could not be reached
because of a loss of line-of-sight in the vicinity of the CT table, despite repositioning of the camera.




       Figure 2: The average percentage of agreement in questionnaire regarding navigated 4-DOF aiming device.
The questionnaire was completed by the three radiologists after completion of the experiments. All users strongly
agreed on the benefit of the aligning aiming device with optical navigation system (Figure 2). Conflictingly, users ex-
pressed dissatisfaction with passive depth control during needle insertion, further details regarding active and passive
depth control are available in [9].

4        Discussion

The usability of a navigated 4-DOF aiming device was confirmed experimentally through the performance of 20 needle
ablations in an anthropomorphic phantom in a clinical environment by radiologists of varying levels of experience; 19
of the 20 planned targets were reached. Further usability assessment was performed by radiologists via the completion
of a questionnaire.
The strongly positive attitude towards the navigated aiming device may be related to a number of factors: questionnaire
participants strongly agreed that the learning curve for maneuvering the device was flat, as well as strongly agreeing
that the needle stabilization achieved was adequate. Furthermore, when questioned about the combination of the naviga-
tion system with the aiming device, all participants strongly agreed that navigation was useful. Navigated needle guid-
ance may provide a number of advantages over existing needle guidance technologies. The lack of spatial orientation
for needle alignment, as well as the requirement for iterative insertion supervised by several control image acquisitions,
are major limitations of existing approaches such as those presented in [6] and [8]. To our knowledge, only one solution
currently provides a similar functionality [7]. This system is, however, limited by a non-automatic registration process
as well as a steeper learning curve.
Respondents indicated that the guidance capabilities of the aiming device may need to be extended by active depth con-
trol, which shows, in real-time, the distance between planned target and current needle tip. Active depth control pro-
vides additional security in case of unexpected patient movements relative to the aiming device or displacement of SM
on the patient skin. Interestingly, in terms of accuracy, active depth control may lead to higher lateral errors at the target
[9], thus its integration into current solution should be carefully considered. Respondents also agreed that automatic
needle alignment, potentially provided by a robotic systems such as [10], could be beneficial (76% agreement). Auto-

                                                            236
matic alignment may help to reduce the time required for needle positioning, a significant drawback of the presented
technique. The time taken for to manually align the needle using an aiming device is a common issue; targeting times
required by SimpleCT [14] and ATLAS aiming device [7] are 15.6 ±16.4 min and 5.03±2.65 min respectively. Further-
more, automatic alignment may be beneficial in the case of multiple parallel needles which could be quickly reached
without the re-positioning of the device, or in the case of thermal ablations carried out under local anesthesia, which re-
quires insertion of both anesthetic needle and treatment applicator exactly at the same position.
Several improvements regarding size of the device and workspace limits were suggested. The major limitation of the
current solution is a narrow range of possible orientation angles (-32°÷32°,-29°÷35°) when compared to other available
solutions [6], [8]; only 30% agreement was achieved when respondents were asked if the angular range of the device
was sufficient. Maximum reachable angles with the SeeStar needle guide is 60° around each plane, while SimpleCT
(NeoRad, Norway), which is a passive laser alignment systems fixed on the wheeled stand near the CT table, can be
oriented at up to 45° in the transversal or the sagittal plane angles respectively. Additionally, the size of the aiming de-
vice (due to the translational movement capabilities) was also considered as a limitation (80% agreement). This is par-
ticularly important for trajectories starting dorsally, potentially leading to a collision of the aiming device with CT table,
as was observed during the experiment, where one trajectory could not be followed.

5        Conclusions

The usability of a novel navigated 4-DOF aiming device was assessed through the completion of a clinically relevant
phantom study and subsequent questionnaire; the system was found to provide both sufficient needle stabilization as
well as a relatively flat learning curve. The device can be integrated into existing IR workflows, potentially improving
the clinical outputs of minimally invasive, percutaneous needle procedures. Further investigations are required to both
improve the angular capabilities of the aiming device and perform a quantitative assessment of achieved targeting accu-
racy and times required for needle alignment.

5        References

[1]    J. R. McCarley and M. C. Soulen, “Percutaneous ablation of hepatic tumors” Seminars in interventional
       radiology, vol. 27, no. 3, pp. 255–60, Sep. 2010.
[2]    J. M. Racadio, D. Babic, R. Homan, J. W. Rampton, M. N. Patel, J. M. Racadio, and N. D. Johnson, “Live 3D
       guidance in the interventional radiology suite.,” AJR. American journal of roentgenology, 189 (6), Dec. 2007.
[3]    P. Schullian, G. Widmann, T. B. Lang, M. Knoflach, and R. Bale, “Accuracy and diagnostic yield of CT-guided
       stereotactic liver biopsy of primary and secondary liver tumors,” Computer aided surgery, vol. 16, no. 4, 2011.
[4]    C. K. Narsule, R. Sales Dos Santos, A. Gupta, M. I. Ebright, R. Rivas, B. D. T. Daly, and H. C. Fernando, “The
       efficacy of electromagnetic navigation to assist with computed tomography-guided percutaneous thermal
       ablation of lung tumors.,” Innovations (Philadelphia, Pa.), vol. 7, no. 3, pp. 187–90, 2012.
[5]    R. Khadem, C. C. Yeh, M. Sadeghi-Tehrani, M. R. Bax, J. a Johnson, J. N. Welch, E. P. Wilkinson, and R.
       Shahidi, “Comparative tracking error analysis of five different optical tracking systems.,” Computer aided
       surgery�: official journal of the International Society for Computer Aided Surgery, vol. 5, no. 2, pp. 98–107,
       Jan. 2000.
[6]    A. Magnusson, E. Radecka, M. Lönnemark, and H. Raland, “Computed-Tomography-Guided Punctures Using a
       New Guidance Device,” Acta Radiologica, vol. 46, no. 5, pp. 505–509, Jan. 2005.
[7]    G. Widmann, P. Schullian, M. Ortler, and R. Bale, “Frameless stereotactic targeting devices�: technical
       features, targeting errors and clinical results,” Int J Med Robotics Comput Assist Surg, vol. 8, no. 1, pp. 1–16,
       2012.
[8]    K. Brabrand, T. M. AaløKKen, G. a. Krombach, R. W. Günther, R. Tariq, a. Magnusson, and P. G. Lindgren,
       “Multicenter Evaluation of a New Laser Guidance System for Computed Tomography Intervention,” Acta
       Radiologica, vol. 45, no. 3, pp. 308–312, Jan. 2004.
[9]    D. Wallach, S. Weber, R. Bale, and G. Widmann, “Comparison of freehand-navigated and aiming device-
       navigated targeting of liver lesions,” International Journal of Medical Robotics and Computer Assisted Surgery,
       Feb, 2013.
[10]   G. Kronreif, J. Kettenbach, M. Fürst, M. Kornfeld, W. Ptacek, and M. Vogele, “IntraROB—a programmable
       targeting device for interventional radiology,” International Congress Series, vol. 1281, pp. 407–411, May 2005.
[11]   T. Oliveira-Santos, B. Klaeser, T. Weitzel, T. Krause, L.-P. Nolte, M. Peterhans, and S. Weber, “A navigation
       system for percutaneous needle interventions based on PET/CT images: Design, workflow and error analysis of
       soft tissue and bone punctures.,” Computer aided surgery, vol. 16, no. 5, pp. 203–19, Jan. 2011.
[12]   G. Toporek, D. Wallach, S. Weber, R. Bale, and G. Widmann, “Cone-beam Computed Tomography-guided
       Stereotactic Liver Punctures: A Phantom Study.,” Cardiovascular and interventional radiology, May 2013.

                                                             237
[13]   T. Oliveira-santos, M. Peterhans, S. Hofmann, and S. Weber, “Passive Single Marker Tracking for Organ
       Motion and Deformation Detection in Open Liver Surgery,” Proceedings of the Second international conference
       on Information processing in computer-assisted interventions, pp. 156–167, 2011.
[14]   C. Moser, J. Becker, M. Deli, M. Busch, M. Boehme, and D. H. W. Groenemeyer, “A novel Laser Navigation
       System reduces radiation exposure and improves accuracy and workflow of CT-guided spinal interventions: a
       prospective, randomized, controlled, clinical trial in comparison to conventional freehand puncture.,” European
       journal of radiology, vol. 82, no. 4, pp. 627–32, Apr. 2013.




                                                         238