=Paper= {{Paper |id=Vol-1498/HAICTA_2015_paper51 |storemode=property |title=The Front End Design of a Health Monitoring System |pdfUrl=https://ceur-ws.org/Vol-1498/HAICTA_2015_paper51.pdf |volume=Vol-1498 |dblpUrl=https://dblp.org/rec/conf/haicta/PetrellisBG15 }} ==The Front End Design of a Health Monitoring System== https://ceur-ws.org/Vol-1498/HAICTA_2015_paper51.pdf
  The Front End Design of a Health Monitoring System

                 Nikos Petrellis1, Michael Birbas2, Fotios Gioulekas3
  1
    Department of Computer Science and Engineering, Technology Educational Institute of
                        Thessaly, Greece, e-mail: npetrellis@teilar.gr
 2
   Analogies S.A. and Electrical Engineering and Computer Technology Dept., University of
                     Patras, Greece, e-mail: michael.birbas@analogies.eu
           3
             Sub-directorate of Informatics, University General Hospital of Larissa
                          Larissa, Greece, e-mail: gioulekas@uhl.gr


       Abstract. In this paper an efficient e-health platform based on a low-cost
       sensor controller system is presented, exhibiting enhanced key characteristics
       able to provide broad coverage of medical scenarios in a reliable and flexible
       way. The heart of the system is a low-cost sensor controller capable of
       performing both simple medical tests and more advanced ones communicating
       with a Gateway and a tablet or smart phone providing instructions to the
       patient. Equipped with a simple and flexible communication protocol for data
       and command exchange, the developed platform is capable of readily
       supporting a variety of sensors with different sampling profiles. Furthermore,
       first promising results of on-going work pave the way for achieving
       considerable enhancement of sensors’ accuracy (close to high-cost commercial
       ones) and significant extension of platform’s portability through power
       consumption minimization. These characteristics have been verified by
       experimenting with various medical scenarios one of which is demonstrated
       here in detail.


      Keywords: health monitor, sensors, communication protocol, low power




1 Introduction

Modern mobile health monitoring systems are classified to two major categories, a)
systems based on commercial sensor infrastructures, b) systems based on low-cost
open source solutions. Specifically, in (Gay and Leijdekkers, 2007), the authors have
proposed a monitoring system based on wireless sensors and smart phones to monitor
the wellbeing of high risk cardiac patients exploiting the usage of a real-time ECG.
They use a Bluetooth enabled blood pressure monitor and weight scale from A&D
Medical A&D (Medical website, 2007) and an integrated Bluetooth
ECG/Accelerometer sensor from Alive Technologies (Alive Technologies, 2007).
These sensors communicate with the smart-phone using Bluetooth technology. The
smart-phone connects to WiFi or through GSM and transmits data to the remote
gateway. The sensors are based on commercial products and it requires the usage of
an expensive smart-phone to serve as first hot-spot before the connection to the
server. Additionally, the work in (Chan et al, 2008) proposes also a similar




                                           426
monitoring architecture like that in (Gay and Leijdekkers, 2007) based on
commercial sensors.
   Furthermore, the work proposed in (Mukherjee et al, 2014) introduces a
perception layer, where a processing device incorporates Bluetooth, Zigbee και
WiMAX protocols to connect the patient sensors to the monitoring infrastructure and
collect the corresponding data. These sensors produce raw values of data which are
wirelessly relayed to a central transceiver unit worn by the patient. This transceiver
unit processes the raw data and converts it into meaningful metadata. The authors of
this work focus on the protocol implementation rather than low-complexity and low-
consumption sensors and integration platform. Authors in (Patel et al, 2012) analyze
and review the available sensor technology, communication technology, and data
analysis techniques towards the implementation of wearable systems and patient
monitoring, denoting the importance of the development and usage of both high-
accurate and low-energy sensors. It is also suggested in (Khelil et al, 2014) an
experimental Body Sensor Network platform that measures important health
indications, and maps the health status of the individual carrying the platform into a
set of predefined classes. They employ a commercial-off-the-shelf (COTS) low-cost
platform, the e-Health Sensor Shield (Cooking-Hacks, 2013) from Libelium Cooking
Hacks which relies on two popular sensing and computing platforms i.e., Arduino
and Raspberry. However, no information is given for the important features of such a
platform, like the communication protocol employed for data/command/exchange,
power consumption and sensors’ accuracy. Finally, the work proposed in (Granados
et al, 2014) is based on a gateway design that performs digital filtering of vital signs
signals, as well as streaming to Web clients. The powering capabilities of the
gateway are based on PoE (power over Ethernet) that can also serve as a power and
data source for wired sensors and building automation appliances. Alternatively,
there exist a number of commercial platforms like e-Shimmer Health BSN (The
Shimmer Platform, 2008), and the Simband health sensor platform (SIMBAND,
2014) from Samsung which are expensive solutions though with proprietary
components without providing HW open-source features. The MIThril system
(MIThril, 2003) is based on monitoring using wearable systems that employ multiple
sensors that are typically integrated into a sensor network either limited to bodyworn
sensors or integrating body-worn sensors and ambient sensors. However, such
systems by design are not suitable for long-term health monitoring.
   As far as the current work is concerned, the low-cost e-Health kit of Libelium-
Cooking Hacks (Cooking-Hacks, 2013) has been employed as basis (sensor
controller) for the implementation of an efficient e-health platform. In comparison
with similar works we further focused in significantly improving its key features,
having developed a communication protocol able to serve any kind of sensors’
sampling profiles and derived first promising results in sensor’s accuracy
enhancement and power consumption minimization, aiming towards the development
of a low cost flexible e-health sensor platform able to support in a reliable as many
medical cases as possible.
   The description of the work performed and the developed platform are organized
as follows: in Section 2 a description of the overall e-health platform is given while
in Section 3 we focus on the sensor controller. In Section 4 the developed
communication protocol between the Gateway and the Sensor Controller is




                                           427
described, in Section 5 first promising results of on- going work are presented while
finally in Section 6 a representative medical scenario is described demonstrating the
effectiveness of the presented techniques and methods.


2 System Description




Fig. 1. The framework of the health monitoring system

   The system description of the heath monitoring system is shown in Fig. 1. In such
a system, the monitored patient, accepts instructions through a smart device (phone
or tablet) offering a comprehensive user interface like e.g. a wide touch screen with
large buttons or a mobile phone/tablet with voice recognition system
   Medical sensors employed in this use include both simple tests (temperature,
respiratory measurements, skin conductance, etc) and/or more advanced ones (like
Electrocardiograms-ECG, Electromyograms-EMG) which are connected to the
Sensor Controller usually based on a lightweight (but powerful enough to cover
system specifications) portable microcontroller module. A Gateway is furthermore
employed which is in most cases positioned at patient’s personal place and is
connected to the Sensor Controller through a wireless protocol like Wi-Fi.
   A communication protocol is responsible for the transfer of the Gateway
instructions to the Sensor Controller concerning the sensors that should be used and
their sampling frequency. It should be stressed that the choice of an appropriate
communication protocol is not a trivial issue since a variety of sensors has to be
controlled with different sampling profiles.
   Power consumption is another crucial issue since it is directly aligned with the
degree of the platform portability that could be provided to the patient. To ensure that
portability is supported, the system power supply, usually a light battery, should last
sufficiently long before being recharged. It should be noted at this point that the main
parameters through which power consumption can be affected are limited to the




                                           428
platform’s microcontroller operation and the wireless communication protocol
controlled, since the power required by the controlled sensors cannot be touched.
   The Gateway is providing an intermediate node between the patient, the
doctor/hospital and the tablet/smartphone. Its main tasks include the communication
with the Sensor Controller and the tablet to forward the instructions to the patient and
to gather the medical test results from the Sensor Controller. These test results can be
forwarded either directly to the doctor or a hospital data base or to the cloud. Some
trivial tasks and decisions can be directly taken by the Gateway like e.g. in the case
where the room temperature exceeds predefined limits, while it should also activate
appropriate alarms if the patient delays to respond, etc.
   The ordinary sensors used in health monitoring systems are grouped in two
categories: simple sensors with analog interface (like temperature, respiratory, etc)
which might call for some special range adaptation circuitry to connect to
corresponding ADC channels and smart sensors with digital interface (like blood
pressure measurement, etc) with predetermined accuracy.


3 Sensor Controller




Fig. 2. Block Diagram of the Sensor Controller

   As shown in Fig. 2, the Sensor Controller is a “mobile microcontroller board the
rough functionality of which is described in the block diagram above. The analog
sensors are connected to separate ADCs and/ or to different channels of the same
ADC. So as to achieve matching of the sensor output to the ADC input, without
linearity degradation, range adaptation circuits might have to be employed, including
either passive components like voltage dividers or Wheatstone bridges in the simplest
case or active circuits like operational amplifiers, while voltage shifting circuits may
also be required. Such range adaptation circuits have to be designed carefully so as to
ensure high linearity and low noise.
    As far as smart sensors with digital interface are concerned (Fig2), they are
connected to serial buses like Serial Peripheral Interface (SPI), I2C or
USART/UART according to the interface provided by the specific sensor. However,
the communication boards used to transfer information between the Sensor
Controller and the Gateway may also need to be controlled by one of these interfaces.
This might lead to conflicts regarding the sensors that should be connected directly to




                                             429
the Sensor Controller since the serial bus interface resources are limited. The sensors
with the simplest communication protocol can be alternatively connected to General
Purpose I/O (GPIO) pins that will be configured to operate in the same way as the
serial bus supported by the sensor. However, the software drivers that will be needed
to emulate the operation of such a serial bus may be a significant overhead.
    It is important that the Processing Unit (PU) and the size of the internal RAM and
Flash memory allow for the use of a Real Time Operating System (RTOS) so that
several real-time tasks could be handled in parallel. Furthermore, if the RTOS offers
a Graphical User Interface (GUI), the development and debugging process of the
health monitoring system is also simplified reducing the time to market. Another
important characteristic is the size of the SRAM and Flash memory in the sense that
it should be adequate to store firmware with high complexity. As far as real time
operation is concerned, the Sensor Controller is usually placed on a table or carried
by the patient when the medical tests have to be performed. The Sensor Controller
communicates with a Gateway through a wireless protocol like Wi-Fi or Bluetooth.
    The current work uses as reference development platform the low cost e-Health kit
of Libelium-Cooking Hacks, which exhibits most of the aforementioned features.
The supported sensors with analog interface can measure: temperature, skin
conductivity, position, breathing airflow, ECG and EMG (EMG cannot operate
concurrently with ECG) while the digital sensors offered are: blood pressure
(connected to UART), measurement of pulse and oxygen in blood (requires 8 GPIO
pins) and glucometer (connected also to UART and thus cannot operate with blood
pressure sensor concurrently). The e-Health kit platform can be directly plugged onto
an Arduino development board or to a Raspberry PI through an Arduino bridge. The
basic functionality offered by the Arduino bridge is the rearrangement of the
Raspberry digital pins so as to provide Arduino-compatible headers. An additional
ADC has been also placed on the Arduino bridge to increase the number of analog
inputs offered by the Raspberry PI.
    Due to the higher processing power (32-bit ARM core) of Raspberry PI and its
increased RAM/ROM size compared to the 8-bit AVR microcontroller that forms the
core of Arduino, Raspberry PI was selected to serve as Sensor Controller. Moreover,
the advanced GUI offered by the Raspbian OS installed on the Raspberry PI board
makes possible the use of the Sensor Controller in stand-alone mode. An SD card is
required by the Raspberry PI to install the Raspbian OS which is also used to store
temporary data and command files required for the communication with the
Gateway. Finally, the stacking of the e-Health kit over the Arduino bridge and the
Raspberry PI allows for the placement of a wireless communication module between
the Arduino bridge and the eHealth kit as shown in Fig. 3. Both Bluetooth and Wi-Fi
interfaces were considered as candidates for the wireless communication between the
Sensor Controller and the Gateway. The Wi-Fi interface was finally selected due to
its higher reliability and the incompatibilities discovered between the Bluetooth card
used at the Sensor Controller and the smart phones tested. The connectors on the top
of Fig. 3 are used to connect the sensors listed above.
    Based on this experimental setup (Fig. 3), we focused on the definition and
implementation of an appropriate communication protocol for the health monitoring
system while also successful results were derived so far on both the improvement of
sensors’ accuracy and minimization of power consumption as part of on-going work.




                                         430
4 Communication Protocol

   The communication protocol is responsible for the data and command exchange
between the Gateway and the Sensor Controller. The commands are used by the
Sensor Controller to schedule the appropriate intervals that the various sensors
should be sampled according to the supported medical case study. Various methods
like HTML server or clouds could have been employed to support the wireless data
exchange between the Gateway and the Sensor Controller, but in our case we
preferred to have the Gateway configured as an FTP server since exchanging
command and data text files is more convenient for debugging and logging purposes.
This message passing communication protocol is implemented through two file
types: command and data, that are exchanged between the Gateway and the Sensor
Controller. The command file has the format shown in the example of Fig. 4 and is
prepared by the Gateway according to instructions set by the supervisor doctor.
                        	
   	
  


                                    WiFi	
                        eHealth	
  
                                    module	
                      kit	
  




                                    Arduino	
                Raspberr
                                    bridge	
                 y	
  PI	
  
                                                             model	
  B	
  

Fig. 3. Stacking of the basic modules of the Sensor Controller                  	
  


   The first line of such a command file defines whether the system will go to sleep
mode for a predefined interval or will perform a sampling scenario (lines starting
with -- are comments). In the case of Fig. 4, the first line determines that a sampling
scenario is to take place. Specifically, in the first line (of Fig. 4), the first character is
either (R) or (S) denoting that the sensor controller will either enter a sampling or a
sleep mode respectively. Two integer numbers follow which represent parameters
Q(uantum) and I(nterval). The Q parameter represents a time interval expressed in
us, while the I parameter defines that the command file will be read again by the
Sensor Controller after a delay of I ms. Ιf the application software on the side of the
Gateway does not change this file, the same sampling scenario will be repeated by
the Sensor Controller (every I ms).
   Each pair of the following lines (ignoring the comments) is assigned to a specific
sensor. The first line of each pair of lines associated with a specific sensor defines
only an integer number N that corresponds to a sampling interval equal to N×Q (in
ms) for this sensor. If N is selected to be 0 this means that the specific sensor is not
used. The second line in each pair is reserved for future options.
   Based on the specification described above, the Sensor Controller software
initially reads the command file and starts a loop during which, the sensors with non-




                                                  431
zero N are sampled. A sleep interval of Q ms is inserted before the next iteration.
When each sensor s is sampled it is omitted from the sampling process for the next
Ns iterations in order to create the Ns×Q pause interval between successive samples.
   The data samples are stored in files that are transferred back to the FTP server.
The data files can contain mixed sensor indications distinguished by the sensor name
and followed by the sampled value and a timestamp. Optionally, different files can be
used to separate the sensor values according to the configuration set by the Gateway.




Fig 4. Sampling ECG (every 100ms) and patient position (every 1 sec) sensors.




5 On Going Work

   The on-going work is towards enhancing the sensors’ accuracy and lowering
significantly the platform’s power consumption.
Sensor Accuracy enhancement
   As already described in section 3, for specific sensors (like analog ones) it is
necessary that they undergo a (hardware) calibration scheme (e.g. use of Wheatstone
bridge techniques etc) to update default parameters so as to ensure sufficient
accuracy. These (hardware) techniques suffer however from several drawbacks, such
as sensitivity to environmental, temperature and power supply variations, etc.
   In contrast to this approach on going work has been focusing in the improvement
of the accuracy of this type of sensors measurements through software instead of
modifying the hardware of the eHealth development platform. More specifically, the
calibration values are stored within configuration files that are updated by the
Raspberry PI microcontroller system using information about environmental
conditions. Then, a filtering procedure takes place using a simple moving average
with extreme value exclusion. More specifically, first a moving average is applied on
the last k sample values (vt, vt-1,…, vt-k+1),and an initial average Ain is estimated.
Following this, each vi of the k values is compared with Ain and if their difference
exceed a predetermined threshold Th, then a corresponding binary flag mi is set to 0
otherwise it is set to 1. After updating all the mi values, the final average Afin is




                                            432
estimated and finally used. Application of this technique to a number of sensors (like
airflow, skin conductance and temperature measurement sensors) has so far given
very satisfactory results with the processed measurements being close enough to ones
from (expensive) medically certified sensors.
Power Consumption
   Another crucial factor where we are targeting to, (on going work), is the
minimization of the power consumption and especially that of the Sensor Controller
(as well as of the WiFi module) through the exploitation of the idle intervals of these
devices between measurements by putting them in a sleep like mode. Experimental
measurements show that the power consumption at full operational mode (when all
the sensors are connected) exceeds 1A which means that if an external 1000mAh
battery is used it would have to be recharged every hour. Sleep mode for the WiFi
module (Roving Networks RN-171) can be readily enabled where only 4uA is
consumed (compared to the 35 mA and the 185-210mA required at normal mode at
the receive/transmit path of the device respectively).
                     	
     Battery	
  Charger	
                                                    USB	
  compatible	
  
                                                                                                    Battery	
  
                            220V	
   	
  	
  USB	
  
                            5V	
                                                                    In	
       	
            Out	
  

                                        Arduino	
  Uno	
  
                                                                                                                      Breadboard	
  
                                                    	
                            5V	
  in	
  pin	
  
                                          Output	
  pin	
  (5V	
  level)	
                                            	
  
                                        USB	
  	
        	
  	
  	
  	
  	
  	
  Output	
  pin	
  



                                        Raspberry	
  PI+eHealth	
  
                                                         	
                  ECG	
  
                                                                        EMG	
  
                                        USB	
  	
   	
          	
  	
  	
  
                                        COM	
  
                                                    Power	
  In	
  (USB)	
  
Fig. 5. Arduino-Rapsberry block diagram to implement low-power consumption

   However this is not the case for the specific (Rapsberry PI) microcontroller board
that does not support sleep or any other low power mode. For this reason, an
additional low cost/power microcontroller that supports sleep mode is employed,
(Arduino in our case) just to turn off and on the Raspberry PI module, as shown in
Fig. 5 through the use of mechanical relays. Specifically, although the Arduino
module is constantly powered by the external battery, the Raspberry PI is powered
through a relay the normal position of which is off. When the Arduino is powered up,
it immediately switches on the relay, powering the Raspberry PI, while also the
Arduino can turn off the Raspberry PI for a predetermined sleep interval if it is
instructed to do so by Raspberry PI. It should be noted that the 8-bit ATmega1280
microcontroller used on the Arduino board draws less than 1mA in normal operation
and less than 1uA in power down mode. Important power savings have been
achieved, and the battery life has been extended from just few hours to days.




                                                                                              433
6 Case Study

    As a demonstration of the issues discussed in previous let’s assume that the
following medical scenario needs to be performed: a) initially the patient position is
tested every second for 4min in order to verify that he has taken the right position for
the following tests, b) an ECG test should be performed for 2min with 0.1sec
intervals, c) after 10 minutes break its blood pressure should be measured just once.
It should be noted here that the patient position has to be concurrently monitored in
both (b) and (c) tests to verify the validity of the examinations and thus the patient
position sensor is sampled every second. To implement this medical scenario, the
following actions have to take place:
    A command file with a sleep instruction may have been deposited at the FTP
server from the last measurement and the Raspberry PI wakes up, reads this file and
falls in sleep mode again until a different than sleep command is found in this file.
The tablet instructs the patient to wear the position and the ECG sensors.
    The patient wears the patient position sensor and the ECG electrodes, switches on
the Sensor Controller module and notifies the tablet that he is ready. The Gateway is
notified through the tablet to change the command file into a new one that instructs
the Sensor Controller to wake up and enter the sampling mode (1st character of first
line of command file is R followed by Q=1000 and I=0)) where only the patient
position sensor is instructed to be sampled every 1x1000mss=1 sec (N=1 in the line
of the command file corresponding to the specific sensor) as well as the EMG sensor
(dummy sampling every 4 min-N=240 in the command file line corresponding to this
sensor). This command file has as effect that the patient position sensor is sampled
immediately (after the command file is read by the Sensor Controller) every second
for N×Q=240x1000=4min when a dummy read to the EMG sensor is
performedThen, immediately the command file will be read again from the FTP
server and the patient position data will be sent to gateway
    The Gateway during the 4 min interval described above prepares the command file
shown in Fig. 4, which instructs the sensor controller to perform the ECG test for
2min. If this modified command file (Fig.4) was not prepared within the 4min
interval, the patient position will be monitored for another 4 min interval before ECG
starts. After the ECG test is completed, the data are sent to the FTP server and the
command file will be immediately read again. This command file could have been
modified by the Gateway in the last 2min interval during the 10 min sleep mode so
that preparations for the blood pressure test take place. The patient in this time
window removes ECG, measures his blood pressure and connects the instrument to
the Sensor Controller.
    After connecting the blood pressure instrument to the Sensor Controller the
patient notifies the tablet that he has performed the measurement. The Gateway will
alter the command file to prepare the transfer of the measurement which takes place
after the Sensor Controller wakes from the 10 min interval. Following the reception
of the blood pressure measurement by the Gateway, the command file can be
changed to sleep mode until the patient is instructed by the tablet to turn it off.
    Regarding the amount of power consumption required, it is derived that the whole
energy consumed per day for this case study 111566.6mA×sec. This means that a




                                          434
1000mAh battery would last 1000×3600/111566.6, i.e. more than 32 days before a
recharge is needed.


7 Conclusions

   A low-cost platform for remote heath monitoring of patients was presented. The
system’s sensor controller is based on the e-health kit from Cookie-Hacks being
capable of performing a number of medical tests from temperature measurement to
Electrocardiograms. A flexible communication protocol allows for the determination
of any type of short or long term sampling of sensors at the desired frequency. This
feature in combination with first promising results derived on sensor’s accuracy
enhancement and power consumption optimization opens the way for a low-cost
system with its key characteristics comparable to those of high cost commercial ones.

Acknowledgments. This work was performed as part of ELTAB project (465435).


References

1. Alive Technologies (2007) [Online]. Available at: http://www.alivetec.com
   [Accessed 03/10/2014]
2. Chan, V., Ray, P. and Parameswaran, N., (2008), Mobile e-Health monitoring: an
   agent-based approach, Communications, IET , Telemedicine and E-Health
   Communication Systems, 2(2), p. 223-230.
3. Cooking-Hacks.(2013)      [Online].   Available     at:    http://www.cooking-
   hacks.com/documentation/tutorials/ehealth-biometric-sensor-platform-arduino-
   raspberry-pi-medical [Accessed 03/20/2014]
4. Gay, V., and Leijdekkers, P. (2007) A Health Monitoring System Using Smart
   Phones and Wearable Sensors, Int. Journal of ARM, 8(2), p. 29-35.
5. Granados, J.; Rahmani, A.M.; Nikander, P., Liljeberg, P. and Tenhunen, H.,
   (2014), Towards energy-efficient HealthCare: An Internet-of-Things architecture
   using intelligent gateways, 4th Int.Conf. MobiHealth, Nov14, Athens, Greece.
6. Khelil, A., Shaikh, F.K., Sheikh, A.A., Felemban, E., and Bojan, H. (2014),
   DigiAID: A Wearable Health Platform for Automated Self-tagging in Emergency
   Cases. 4th MobiHealth, Nov14, Athens, Greece.
7. Medical website (2007) [Online]. Available at: http://www.andmedical.com.au/
   [Accessed: 02/12/2014]
8. MIThril     (2003),     [Online].       Available     at:             http://www.
   media.mit.edu/wearables/mithril/, [Accessed: 02/12/2014]
9. Mukherjee, S., Dolui, K. and Kanti Datta, S.(2014), Patient Health Management
   System using e-Health Monitoring Architecture, Advanced             Computing
   Conference (IACC‘ 14), IEEE International, 21-22 Feb. 2014, p.400 – 405




                                          435
10. Patel, S., Park, H., Bonato, P., Chan, L., and Rodgers, M. (2012) A review of
    wearable sensors and systems with application in rehabilitation. Journal of Neuro
    Engineering and Rehabilitation p. 9-21.