|title=A novel algorithm to predict knee angle from EMG signals for controlling a lower limb exoskeleton 
|pdfUrl=https://ceur-ws.org/Vol-1638/Paper66.pdf
|volume=Vol-1638
|authors=Inderjeet Singh Dhindsa,Ravinder Agarwal,Hardeep Singh Ryait
}}
==A novel algorithm to predict knee angle from EMG signals for controlling a lower limb exoskeleton ==
Mathematical Modeling
A NOVEL ALGORITHM TO PREDICT KNEE ANGLE
   FROM EMG SIGNALS FOR CONTROLLING A
        LOWER LIMB EXOSKELETON
         Inderjeet Singh Dhindsa1, Ravinder Agarwal1, Hardeep Singh Ryait2
                        1
                          EIED, Thapar University Patiala, Punjab, India
                2
                    ECE Department, BBSBEC, Fatehgarh Sahib, Punjab, India
       Abstract. This paper presents an algorithm for estimation of the intended knee
       joint angle from sEMG signals acquired from four muscles of upper limb. The
       algorithm was evaluated by experiments showing the calculated intended mo-
       tion while performing a simple daily life activity of sitting in a squat position
       and standing from a squat position. The proposed algorithm uses Mean Abso-
       lute Value (MAV) and Root Mean square (RMS) as features and a multi-layer
       Back Propagation Neural Network (BPN) for predicting the knee angle. Pro-
       posed algorithm along with the experimental results are presented. The predict-
       ed knee angle can be used to control a lower limb exoskeleton.
       Keywords: Back propagation, Exoskeleton, feature extraction, sEMG, Neural
       Network.
       Citation: Dhindsa IS, Agarwal R, Ryait HS. A novel algorithm to predict knee
       angle from EMG signals for controlling a lower limb exoskeleton. CEUR
       Workshop Proceedings, 2016; 1638: 536-541. DOI: 10.18287/1613-0073-2016-
       1638-536-541
1      Introduction
Exoskeletons for human performance enhancement are wearable devices and ma-
chines that can increase the speed, strength, and endurance of human being. The user
provides control signals for the exoskeleton, while the exoskeleton actuators provide
majority of the power necessary for performing the task. The human operator applies
a scaled down force compared with the load carried by the exoskeleton.
The earliest recorded work in the field of exoskeleton is that of Nicholas Yagn a Rus-
sian army engineer. He was granted a US Patent for his work on “Apparatus for facili-
tating walking, running, and jumping” [1]. Although Yagn's mechanism was designed
to augment running, there is no record that the device was ever built and successfully
demonstrated. Prominent earliest proposed exoskeletons were General Electric’s
“HARDIman” ( Human Augmentation Research and Development Investigation) and
Los Alamos National Laboratory’s“PITMAN”(Powered suit of armour for infantry-
men) [3].
Information Technology and Nanotechnology (ITNT-2016)                                      536
Mathematical Modeling                            Dhindsa IS, Agarwal R, Ryait HS. A novel…
The development in the field of exoskeleton picked pace in 2001 when the U.S. De-
fense Advanced Research Projects Agency (DARPA) started a 50 million project
named “Exoskeletons for Human Performance Augmentation (EHPA)”. The main
goal of EHPA was “to increase the capabilities of ground soldiers beyond that of a
human”. The EHPA program made a number of institutes made research progress in
the technologies related to Exoskeletons.
                        Fig. 1. Block diagram representation of system
Prominent exoskeleton developed under the DARPA project were that of Massachu-
setts Institute of Technology (MIT) [4], Berkley University's “Berkley Lower Extrem-
ity Exoskeleton (BLEEX)” [5] and Sarcos Research Corporation's “Wearable Ener-
getically Autonomous Robot (WEAR)” [6]. All these exoskeletons had one common
feature that human motion intent was measured from force sensor placed between the
human operator and exoskeleton. The exoskeleton was triggered either by the kine-
matic position command or by the dynamic contact force command. This introduced a
delay in the system. This delay was overcome in the next generation of exoskeleton
based on biosensors.
Many of the new proposed exoskeleton predicted the motion intent from EMG sig-
nals. Prominent among these exoskeletons was an exoskeleton named Hybrid Assis-
tive Limb (HAL) developed at the University of Tsukuba, Japan, targeted for both
performance augmenting and rehabilitative purposes [7]. The main advantage of
EMG based exoskeletons is the ability to estimate the forces that will be generated by
the muscles before the actual occurrence of mechanical contractions. This information
is fed into the exoskeleton system such that by the time the physiological muscles
contract the exoskeleton amplifies the joint moment by a preselected gain factor, re-
sulting in a decrease in the reaction time of the human/machine system.
This paper presents an algorithm for estimation of the intended knee joint angle from
sEMG signals acquired from four muscles of lower limb namely Vastus Lateralis
(VL), Semitendinosus (STEN), Biceps Femoris (BF) and Rectus Femoris (RF). The
proposed algorithm uses mean absolute value (MAV) and root mean square (RMS)
Information Technology and Nanotechnology (ITNT-2016)                                 537
Mathematical Modeling                            Dhindsa IS, Agarwal R, Ryait HS. A novel…
for feature extraction and a multi-layer back propagation neural network (BPN) for
predicting the knee angle. The algorithm was evaluated by experiments showing the
calculated intended motion while performing a simple daily life activity of sitting in a
squat position and standing from a squat position. The algorithm and the experimental
results are both presented. The predicted knee angle can be used to control a lower
limb exoskeleton.
                        Table 1. Sensor Position of muscles considered
S.no             Muscle                                     Sensor Position
  1        Vastus Lateralis (VL)             The electrodes need to be placed at 2/3 on
                                          the line from the anterior spina iliaca superior
                                          to the lateral side of the patella
    2      Semitendinosus (STEN)             Electrodes need to be placed at 50% on the
                                          line between the ischial tuberosity and the me-
                                          dial epycondyle of the tibia
    3      Biceps Femoris (BF)               The electrodes need to be placed at 50% on
                                          the line between the ischial tuberosity and the
                                          lateral epicondyle of the tibia
    4      Rectus Femoris(RF)                The electrodes need to be placed at 50% on
                                          the line from the anterior spina iliaca superior
                                          to the superior part of the patella
2       Methodology
Five subjects of mean age 31.08 ± 3.15 years, mean weight 81.42 ± 9.5 Kg and mean
height 1.72 ±0. 08 m respectively, were asked to perform simple daily life activity of
sitting in a squat position from standing position and vice versa. sEMG signals were
recorded from four muscles of lower limb namely Vastus Lateralis (VL), Semitendi-
nosus (STEN), Biceps Femoris (BF) and Rectus Femoris(RF). EMG signals were
recorded using BioTrace+ software for the NeXus-10 biofeedback system at a sam-
pling frequency of 2048Hz. SENIAM recommendations were followed for the skin
preparation and sensor locations [8, 9]. Reference electrode was placed on the tibia
bone. The sensor location for the four muscles are tabulated in Table 1. Block dia-
gram of the joint angle prediction is shown in Fig1. Knee angle was measured using a
linear potentiometer based goniometer using NI USB 6009 data acquisition system at
a sampling frequency of 2048 Hz. The raw sEMG signal was band pass filtered from
20Hz to 500Hz. The purpose of the lower frequency limit was to remove DC offset
and motion artifacts. The high frequency limit prevents the aliasing. This was
achieved by using a zero-lag sixth order recursive Butterworth filter. The sEMG sig-
nal recorded from VL muscle along with knee angle for subject 1 are plotted in Fig 2.
The sEMG signal is not fed directly to a neural network, because of the dimensionali-
ty and random characteristics of the signal. Instead features were extracted and fed to
the neural network. In this paper, two time domain features mean absolute value
(MAV) and root mean square value (RMS) had been used.
Information Technology and Nanotechnology (ITNT-2016)                                 538
Mathematical Modeling                                Dhindsa IS, Agarwal R, Ryait HS. A novel…
         1
𝑀𝐴𝑉 = ∑𝑁
       𝑘=1|𝑥𝑘 |                                                                            (1)
        𝑁
             1
𝑅𝑀𝑆 = √ ∑𝑁    2
         𝑘=1 𝑥𝑘                                                                            (2)
             𝑁
Where 𝑥𝑘 is thevoltage of 𝑘 𝑡ℎ sample, N is number of samples in a segment. The
number of samples was set to be 128 for the current study.
Fig. 2. sEMG signal recorded from Vastus lateralis and measured Knee angle while performing
                             a Stand to Squat to Stand operation
The feature vector is fed to a Back propagation network (BPN). A BPN is used since
it learns by example. When Training is finished it returns the output for a particular
input. BPN are ideal for simple pattern recognition and mapping tasks. The architec-
ture of the BPN used is shown in Fig 3. The adaptive learning algorithm of the BPN
neural network used is explained in following equations:
         1
𝐸(𝑛) = (𝜃𝑚 − 𝜃𝑎 )2                                                                         (3)
         2
𝑤𝑞 (𝑛 + 1) = 𝑤𝑞 (𝑛) + ∆𝑤𝑞                                                                  (4)
𝑚𝑞𝑖 (𝑛 + 1) = 𝑚𝑞𝑖 (𝑛) + ∆𝑚𝑞𝑖                                                               (5)
𝜎𝑞𝑖 (𝑛 + 1) = 𝜎𝑞𝑖 (𝑛) + ∆𝜎𝑞𝑖                                                               (6)
                                                        2
                                  (𝑓𝑁𝑁 (𝑖)−𝑚𝑞𝑖 (𝑛))
∆𝑤𝑞 = 𝜂𝑞 (𝜃𝑚 − 𝜃𝑎 )2 𝑒𝑥𝑝 [− ∑𝐼𝑖                             ]                              (7)
                                        𝜎2𝑞𝑖 (𝑛)
                                        2                       2
                        (𝑓𝑁𝑁 (𝑖)−𝑚𝑞𝑖 (𝑛))       (𝑓𝑁𝑁 (𝑖)−𝑚𝑞𝑖 (𝑛))
∆𝑚𝑞𝑖 = 𝜂𝑚 2𝑒𝑥𝑝 [− ∑𝐼𝑖          2 (𝑛)        ]          2 (𝑛)        (𝜃𝑚 − 𝜃𝑎 )𝑤𝑞 (𝑛)       (8)
                              𝜎𝑞𝑖                     𝜎𝑞𝑖
Information Technology and Nanotechnology (ITNT-2016)                                     539
Mathematical Modeling                                  Dhindsa IS, Agarwal R, Ryait HS. A novel…
            Fig. 3. Multi layer neural network used for predicting the knee angle
                                        2                       2
                        (𝑓𝑁𝑁 (𝑖)−𝑚𝑞𝑖 (𝑛))       (𝑓𝑁𝑁 (𝑖)−𝑚𝑞𝑖 (𝑛))
∆𝜎𝑞𝑖 = 𝜂𝜎 2𝑒𝑥𝑝 [− ∑𝐼𝑖         2 (𝑛)         ]          3 (𝑛)        (𝜃𝑚 − 𝜃𝑎 )𝑤𝑞 (𝑛)         (9)
                             𝜎𝑞𝑖                      𝜎𝑞𝑖
Where 𝜂𝑞 , 𝜂𝑚 𝑎𝑛𝑑𝜂𝜎 are the learning constant, 𝜃𝑚 is the knee angle predicted and 𝜃𝑎
is the actual knee angle recorded by goniometer.
3      Results
Subjects were asked to perform daily life activities like normal walking, stand to sit in
a squat and squat to stand operation, standing up and sitting on a chair and ascending
and descending of stairs. sEMG signals were recorded from VL, STEN, BF and
GMED muscles of lower limb. Knee angle was measured from the potentiometer
based goniometer and NI 6009.
The recorded data was processed and time domain features MAV and RMS were
obtained. The BPN was first trained using the 60% of the data and knee angle was
predicted. The predicted knee angle and actual knee angle for Subject 1 are plotted in
Fig 4. The accuracy of the predicted knee angle using the proposed algorithm is
92.4%. The predicted knee angle could be used to control an exoskeleton or other
orthotic or prosthetic device.
Conclusion
The present study proposed an algorithm for predicting knee angle from sEMG sig-
nals recorded from four muscles of human lower limb. The proposed algorithm accu-
rately predicts the knee angle and could be used for controlling the exoskeleton myoe-
lectrically.
Information Technology and Nanotechnology (ITNT-2016)                                       540
Mathematical Modeling                           Dhindsa IS, Agarwal R, Ryait HS. A novel…
                        Fig. 4. The Predicted and measured knee angle
References
 1. Yagn N. Apparatus for facilitating walking, running and jumping, 1890. US Patent 420,
    179. URL: https://www.google.com/patents/ US420179.
 2. Fickand B, Markinson J. Final Report On HARDIMAN-I Prototype For Machine Aug-
    mentation Of Human Strength And Enduance, Tech.Rep., General Electric Company,
    Schenectady, New York, 1971; 12345.
 3. Moore J. L.A.N. Laboratory. PITMAN, a Powered Exoskeletal Suit for the Infantryman,
    Los Alamos National Laboratory, 1986. URL: http://books.google.co.in/books?
    id=stisGwAACAAJ.
 4. Walsh C, Endo K, Herr H. A quasi –passive leg exoskeleton for load-carrying augmenta-
    tion, International Journal of Humanoid Robotics, 2007; 4: 487-506.
 5. Zoss A, Kazerooni H, Chu A. Biomechanical design of the Berkeley lower extremity exo-
    skeleton (BLEEX), Mechatronics. A SME Transactions on IEEE, 2006; 11: 128-138.
 6. Guizzo E, Goldstein H. The rise of the body bots [robotic exoskeletons], Spectrum, IEEE,
    2005; 42: 50-56.
 7. Lee S, Sankai Y. Power assist control for walking aid with HAL-3 based on EMG and im-
    pedance adjustment a round knee joint, 2002; 2: 1499-1504.
 8. SENIAM group. Recommendations for sensor locations on individual muscles, 2013.
    URL: http://seniam.org/sensor_location.htm.
 9. Stegeman D, Hermens H. Standards for surface electromyography: The European project
    surface emg for non-invasive assessment of muscles (seniam), 2007.
10. Olney SJ, Winter DA. Predictions of knee and ankle moments of force in walking from
    emg and kinematic data, Journal of biomechanics, 1985; 18: 9-20.
Information Technology and Nanotechnology (ITNT-2016)                                   541