=Paper= {{Paper |id=Vol-2023/335-339-paper-54 |storemode=property |title=Development of the autocalibration system for the DGFRS spectrometer based on the double-sided silicon strip detectors |pdfUrl=https://ceur-ws.org/Vol-2023/335-339-paper-54.pdf |volume=Vol-2023 |authors=Vladimir Subbotin,Alla Zubareva,Leo Schlattauer,Alexey Voinov }} ==Development of the autocalibration system for the DGFRS spectrometer based on the double-sided silicon strip detectors== https://ceur-ws.org/Vol-2023/335-339-paper-54.pdf
       Proceedings of the XXVI International Symposium on Nuclear Electronics & Computing (NEC’2017)
                             Becici, Budva, Montenegro, September 25 - 29, 2017



  DEVELOPMENT OF THE AUTOCALIBRATION SYSTEM
   FOR THE DGFRS SPECTROMETER BASED ON THE
     DOUBLE-SIDED SILICON STRIP DETECTORS
       V.G. Subbotin1, A.M. Zubareva1, L. Schlattauer1,2 and A.A. Voinov1,a
   1
       Laboratory of Nuclear Reactions, Joint Institute for Nuclear Research, 6 Joliot-Curie, Dubna,
                                    Moscow region, 141980, Russia
       2
           Department of Experimental Physics, Faculty of Science, Palacký University, 17. listopadu
                                1192/12, 771 46 Olomouc, Czech Republic

                                          E-mail: a voinov@jinr.ru


The detection system of the Dubna gas-filled recoil separator (DGFRS) aimed at the studying of the
SHE nuclei and their decay properties has been modernized during last few years. The new set of
multi-strips double-sided silicon detectors (DSSD) in focal plane of DGFRS is applied now instead
of the old array of 12-strips position-sensitive Si detectors. The total amount of measuring
spectroscopic channels of the registering system has increased also up to 224 channels. It leads to
more precise measuring of the energy and coordinate of the implanted nuclei of the SHE into the
focal detectors and of their decay products. It is important to test multi-channel registering system
and perform energy calibration before carrying out of such unique experiments on the synthesis of
new nuclei from the “Island of stability”. This work is devoted to describe the designed method and
produced specific digital module which allows performing an energy calibration for the all 224
individual spectroscopic channels independently. This device provides automatic bypassing of the all
individual channels one after another imitating charge particles incoming to the each strip of detector
array. Energy of the imitating signal can be chosen from the range of 1 MeV up to 150 MeV with
high amplitude linearity and temperature stability.

Keywords: DGFRS, DSSD, registering system, energy calibration, nuclear electronics,
multiplexer, encoder, operational amplifier

                            © 2017 Vladimir G. Subbotin, Alla M. Zubareva, Leo Schlattauer, Alexey A. Voinov




                                                                                                        335
    Proceedings of the XXVI International Symposium on Nuclear Electronics & Computing (NEC’2017)
                          Becici, Budva, Montenegro, September 25 - 29, 2017




1. Introduction
         More than 50 new isotopes of the nuclei with Z = 104-118 from the predicted “Island of
stability” of superheavy elements (SHE) were synthesized for the first time at the Flerov Laboratory
of Nuclear Reactions (JINR, Dubna) using Dubna Gas-Filled Recoil Separator (DGFRS) and heavy
ions cyclotron U-400 during last 15 years [1]. Due to this research the Periodic Table of Elements
was recently filled with six new elements Z=113 (Nh), 114 (Fl), 115 (Mc), 116 (Lv), 117 (Ts) and
118 (Og).
         During these pioneering experiments in FLNR the appropriate detector module had being
used at the DGFRS focal plane which consisted of the array of twelve position-sensitive Si detectors
surrounded by six side Si detectors (Canberra NV) and of the time-of-flight module (two multi-wire
proportional chambers fulfilled with pentane gas at low pressure about 1 Torr). More detailed
descriptions can be found in our previous papers and references therein [2-4].
         Recently, the focal-plane detector array has been modified to increase the position resolution
of recorded signals and subsequently reduce the probability of observing sequences of random events
that mimic decay chains of implanted nuclei. The new detector assemblage consists of 300 μm
thickness double-sided Si strip detector (DSSD) with 48 mm by 128 mm active area in focal plane
and surrounded by 500 μm six single-sided Si detectors 65 mm by 120 mm each (Micron
Semiconductor Ltd.). Focal-plane DSSD has 1-mm wide strips, 48 at the front side and 128 at the
back side, creating over 6000 individual 1×1 mm2 pixels in one Si wafer. Such high pixelization
helps to achieve superior position resolution for recoil-correlated decay sequences reducing potential
random events. This new Si-detector array was designed, assembled, commissioned off-line and
provided by ORNL. The signals from all detectors were processed using MESYTEC linear
preamplifiers [5]. Further, these analog signals were split into two independent measurement
branches by special spectroscopic splitter-amplifier PA32-64 [6] designed by the DGFRS group.
Thus, all detectors’ signals were processed simultaneously by “analog” electronics (TekhInvest Ltd.,
Dubna) similar to those used in previous DGFRS experiments [1], and by digital electronics system
based on XIA Pixie-16 modules [7].
         This new DSSD assembly and two new independent registering systems were successfully
applied recently in 239Pu, 240Pu + 48Ca experiments [8] and 249-251Cf + 48Ca experiment [9] aimed at
the synthesis and the study of the properties of the new 283-285Fl and 295,296Og isotopes. The FWHM
energy resolution of the implantation detector was 34 to 78 keV depending on the strip, while the
summed signals recorded by the side and implantation detectors had an energy resolution of 147 to
263 keV. For getting better energy resolution for side detectors we plan to order new design with
eight strips for every side detector, thus it will cause increasing of the total number of the
spectroscopic channels by 48 ones in comparison with used present detectors design [8].
         The detectors and registering systems should be calibrated prior to start every experiment,
moreover before every finish of the experiment and sometimes during the experiments also. For this
purpose we perform two reactions, natYb(48Ca, 3–5n)215-220Th to calibrate -range scale and
208
   Pb(48Ca,2n)252No – to SF range scale, as it was described in [4]. There are no problem to calibrate
individual PIXIE-16 channel in digital measuring system (based on the DSP and FPGA, see [7]);
every PIXIE-16 channel is working independently. In “analog” system (CAMAC based, TekhInvest
Ltd.) every measuring ADC works in combination with 16-input analog multiplexer for each of 16
strips; only first signal in time from every 16 strips will be processed by corresponding ADC.
         Besides of nuclear reactions, for the linearity test of the scales we can calibrate our systems
by means of spectroscopic pulser module (for example, ORTEC 419) with precise stable preset
amplitudes [10] imitating income of α-particles and SF events to surface of the Si detectors.

2. 16-channel charge-sensitive preamplifier
         Choosing of MPR-64 from Mesytec [5] as multi-channel charge-sensitive preamplifier with
high resolution, linearity in the energy range of signals up to 300 MeV and with long-term stability
are satisfying our needs during the experiments at the DGFRS. There is just one disadvantage of



                                                                                                    336
    Proceedings of the XXVI International Symposium on Nuclear Electronics & Computing (NEC’2017)
                          Becici, Budva, Montenegro, September 25 - 29, 2017



using these devices together with “analog” measuring branch.
Preamplifier MPR-64 has 64 individual channels performed as
four separate PCBs containing 16 channels each. There is also
separate Pulser Input for each PCBs, thus signals from outer
pulser can go just simultaneously to all 16 channels of one
PCB. So, due to this feature of MPR-64 there is no possibility
to perform the energy calibration for the individual channel in
“analog” registering system based on ADP-16 (TekhInvest
Ltd.) which has on-board SAR ADC working in combination
with 16-input multiplexer with priority encoder. To bypass
this inconvenience in calibration of the individual channel in
the “analog” branch we designed new multi-channel charge-
sensitive PA-16 preamplifier (Figure 1), which was recently
tested and presented during the symposium NEC’2015 [6].
       The main characteristics are presented below and in
Table 1:
        Input/Output polarity         - Inverted
       Energy sensitivity (Si)         - 8 mV/MeV
       Noise output performance - See Table 1
       Negative feedback constant - 10MΩ/5.6 pF                  Figure 1. The front panel and PCB
       Power supply              +6V     200 mA                  view of 16-channel charge-
                                                                  sensitive PA-16 preamplifier [6]
                                  -6V     110 mA
         For calibration of the individual channels with external pulse generator, the CMOS low-
voltage 16-channel ADG706 multiplexer [11] was applied operating in de-multiplexing mode. It
switches the input pulse from the precision spectroscopic pulser to one of the 16 outputs with a
number chosen by address code preset by special control BOKAL module.
         To process signals from the
whole detector module (48 ‘front’ Table 1. Noise performance measured for PA-16 using
strips and 128 ‘back’ strips of DSSD; the ORTEC 575A spectroscopy amplifier set at 1.5 μs,
and 48 ‘side’ strips) we need 14 near-Gaussian shaping
preamplifier PA-16 modules. They are
realized as 1M CAMAC modules, and
have LEMO input for pulser and two
DB-37 connectors for 16 input signals
from detectors and 16 – output,
correspondently. At the back of the
PCB of every PA-16 module there are
4 pins going to CAMAC bus for
controlling the number of certain spectroscopic channel in the preamplifier by BOKAL module
providing binary code “1 of 16” through CAMAC bus to the each preamplifier in the crate. The
characteristics of the spectroscopic pulses from PA-16 are presented in Figure 2.




 Figure 2. The shapes of the input signal from precise СГ-219 [12] pulser (in orange) and output
 signal after PA-16 (in bleu) are shown



                                                                                                    337
    Proceedings of the XXVI International Symposium on Nuclear Electronics & Computing (NEC’2017)
                          Becici, Budva, Montenegro, September 25 - 29, 2017



3. Controller for spreading out and counting of the pulses
         New mixed analog-digital Block for Organization of
CALibration (BOKAL) was design to perform testing and
energy calibration “off-line” of the individual measuring
channels of the DGFRS registering system. Impulses of certain
amplitudes (corresponding to the energies of 6, 12, 70 and 140
MeV) from precise spectroscopic СГ-219 generator [12]
calibrated by α–source comes to the input “Гек” of the module
and then are shared into three groups of outputs (128 ПУ (yellow
LEMO) connectors for rear DSSD strips, 48 ПУ ФОК (red) for
front strips and 48 ПУ БОК (blue) for side strips). Every
connector goes to Pulser input of the preamplifier PA-16 serving
corresponding front, rear and side strips of the detector array.
BOKAL module allows calibrate measuring channels both
automatically with pre-defined energy amplitude of the pulser or
in manual mode. In auto-mode pulses start from pressing button
“ПУСК” and will stop after BOKAL will count 1024 (or
2048/4096 – optionally) events for every channel. The
calibration procedure can be stopped also manually by “СТОП”
button. In manual mode user should choose needed block of
preamplifiers by button “+1 № блока” and needed channel by
button “+1 № ПУ”. The start will be after push the button
“ПУСК ГЕК” and will finish after push “СТОП”. The counting
and numbers of the chosen block and preamplifier channel can
be reset by button “СБРОС”. The blue LED lamps indicate
current block in binary code, and green lamps indicate current Figure 3. Front view of the
channel. After reset the counting will begin from the first module module BOKAL (Block for the
and the first channel (with the lamps off). The output signals Organizing of the CALibration)
from BOKAL are shown in Figure 4.




Figure 4. The input СГ-219 [12] pulser signal and output signals going to corresponding PA-16
(front, rear and side) are shown in yellow (1), in blue (2), magenta (3) and green (4)

        The propagation delay of the input signal for BOKAL is 34 ns. The module can process the
spectroscopic signals with rise time starting from 10 ns. The upper speed limit for coming pulses is
100 kHz. The transmission coefficient equals to 1. The output polarity of the signals (for front, rear
and side preamplifiers) can be set by user on-board. The requested power is +6V/-6V and
180mA/100mA. The device can be performed both as CAMAC-standard module or NIM-standard.




                                                                                                    338
    Proceedings of the XXVI International Symposium on Nuclear Electronics & Computing (NEC’2017)
                          Becici, Budva, Montenegro, September 25 - 29, 2017



4. Conclusion
        The unique system for the calibration of the registering system based on the multi-strip
double-sided silicon detectors has been designed and realized by our research team. It allows imitate
incoming of the charge particle to the surface of the DSSD. User can choose appropriate energy from
1 MeV up to 150 MeV with pulser preset. To achieve such functionality we designed 16-channel
spectroscopic charge-sensitive preamplifier and specific logical module (BOKAL) which provides
automatic energy calibration of every channel of the multi-channel measuring system using globally
controlled sequential distribution of spectroscopic pulses. This allows perform calibration both in
manual “channel by channel” mode or in auto mode. The auto mode is important time saver before
every experiment in comparison with using manual calibration by connecting the pulser to each
channel by cable manually. Presented part of the spectrometer also minimizes the human errors
during the calibration process.


5. Acknowledgement
        The current work was carried out with the support of the Russian Foundation for Basic
Research by grants No. 13-02-12052 and No. 16-52-55002; and by Moscow Region Government
through Grant No. 736/36-16.09.2014. Also, this work was partially supported by the internal student
grant IGA of Palacký University in Olomouc, Czech Republic, No. IGA_PrF_2015_017.


References
[1] Oganessian Yu.Ts. and Utyonkov V.K. Super-heavy element research. // Rep. Prog. Phys. 78,
036301 (2015).
[2] Tsyganov Yu.S. et al., Nucl. Instr. and Meth. in Phys. Res. A 525, (2004), pp. 213-216.
[3] Tsyganov Yu S, A. N. Polyakov, Nucl. Instr. and Meth. in Phys. Res. A 513 (2003) 413; A 558
(2006) 329-332; A 573 (2007) 161.
[4] Voinov A.A. et al., Calibration of the silicon position-sensitive detectors using the implanted
reaction products, in Proceedings of the XXIII International Symposium on Nuclear Electronics and
Computing “NEC’2011”, Varna, Bulgaria, 2011, pp.286-291.
[5] Rajabali M.M. et al. // Phys. Rev. C 85, 034326 (2012): Mesytec GmbH & Co. KG, Multichannel
Logarithmic Preamplifier; http://www.mesytec.com.
[6] Subbotin V.G. et al. New analog electronics for the new challenges in synthesis of superheavy
elements // Physics of Particles and Nuclei Letters, ISSN 1547-4771, eISSN: 1531-8567, 2016, vol.
13, No. 5, pp. 557–560.
[7] Grzywacz R. et al., Nucl. Instrum. Methods. Phys. Res., Sect B 261, 1103 (2007);
http://www.xia.com.
[8] Utyonkov V. K. et al. Experiments on the synthesis of super-heavy nuclei 284Fl and 285Fl in the
239,240
        Pu +48Ca reactions // Physical Review C 92, 034609, 2015.
[9] Voinov A.A. et al. Results from the recent study of the 249-251Cf + 48Ca reactions, in Proceedings
of the International Symposium on “Exotic Nuclei EXON-2016”, ISBN 978-981-3226-53-1, Kazan,
Russia, 2016, pp. 219-223.
[10] www.ortec-online.com.
[11] www.analog.com
[12] Subbotin V.G., Iliev S.N., Voinov A.A., JINR Report P13-2004-97, Dubna, 2004.