        <article-title>Smart System for Real-time Monitoring of Vital Parameters</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>Richard Bagín</string-name>
          <email>richard.bagin@savba.sk</email>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Bojana Koteska</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Magdalena Kostoska</string-name>
          <email>magdalena.kostoska@finki.ukim.mk</email>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Nevena Ackovska</string-name>
          <email>nevena.ackovska@finki.ukim.mk</email>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Anastas Mishev</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Ana Madevska Bogdanova</string-name>
          <email>ana.madevska.bogdanova@finki.ukim.mk</email>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Milan Tysler</string-name>
          <email>tysler@savba.sk</email>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <aff id="aff0">
          <label>0</label>
          <institution>Institute of Measurement Science, Slovak Academy of Sciences</institution>
          ,
          <addr-line>Dubravska cesta 9 Bratislava 841 04</addr-line>
          ,
          <country country="SK">Slovakia</country>
        </aff>
        <aff id="aff1">
          <label>1</label>
          <institution>Ss. Cyril and Methodius University, Faculty of Computer Science and Engineering</institution>
          ,
          <addr-line>Rugjer Boshkovikj 16, 1000 Skopje, North</addr-line>
          <country country="MK">Macedonia</country>
        </aff>
      </contrib-group>
      <abstract>
        <p>Wearable real-time systems that collect and intelligently analyze data on respiration, heart rate, oxygen saturation, blood pressure, and body temperature could assist medical personnel in selecting the most appropriate response during highly stressful situations in both military and civilian contexts, such as terrorist attacks, IED incidents, or rescue operations. These systems will issue an alert if there is a change in a person's health status to prevent overlooking critical health changes. We propose designing and developing a prototype of a patch-like device and a methodology that enables the continuous monitoring of vital parameters of personnel or victims. The system involves using Artificial Intelligence to create software capable of real-time prediction of the parameters and alerting.</p>
      </abstract>
      <kwd-group>
        <kwd>eol&gt;vital parameters</kwd>
        <kwd>smart patch</kwd>
        <kwd>artificial intelligence</kwd>
        <kwd>medical device</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec id="sec-1">
      <title>1. Introduction</title>
      <p>
        Wearable real-time physiological status monitors, designed as patch-like devices, have been developed
to gather and analyze vital parameters including respiration rate (RR), heart rate (HR), blood oxygen
saturation (SpO2), electrocardiogram (ECG), blood pressure (BP), and body temperature [
        <xref ref-type="bibr" rid="ref1">1</xref>
        ]. These devices
hold promise for assisting first responders and remote personnel in both military and civilian settings,
especially during urgent situations such as terrorist attacks, IED explosions, or rescue operations.
      </p>
      <p>The ability to instantly analyze an individual’s health status, particularly those in active roles like
rescuers and emergency crews, is vital for making quick decisions and managing crises efectively [ 2]. In
this paper we are describing the prototype of a patch-like sensor to address post-attack activities related
to managing wounded victims, thus increasing their chances of survival. The sensor is intended to be
placed on victims’ chests after initial triage, with a focus on those labeled yellow and green according
to the START (Simple Triage and Rapid Treatment) triage approach [3]. The device would issue alerts
when the medical status changes from green-to-yellow (GtoY) or yellow-to-red (YtoR), thereby elevating
the priority of afected victims.</p>
      <p>Our primary research goal is to develop an integrated sensor system capable of real-time health
monitoring and alerting, enhancing the eficiency and accuracy of emergency triage processes. The
main contribution of this proposed solution is the integration of sensors for monitoring HR, RR, and
other parameters, providing real-time monitoring of the health status of emergency crews or generating
immediate alerts if the patient’s health status deteriorates during the triage process. Additionally, the
integration of Artificial Intelligence (AI) methods, represents a new generation of health monitoring
solutions.</p>
      <p>The successful development of this merged system is expected to significantly impact the readiness
of squads to handle emergencies that may afect crew and victims’ health status in real-time.</p>
      <p>The paper is organized as follows. The Related Work section describes similar patch solutions.
Section 3 provides an overview of the system, including the hardware part, Bluetooth low energy
implementation, and the Android Software application. The conclusion is given in the last section.</p>
    </sec>
    <sec id="sec-2">
      <title>2. Related Work</title>
      <p>In exploring existing research, several studies have addressed the pressing need for continuous patient
monitoring and innovative healthcare solutions. Yeri and Shubhangi [4] highlight the necessity of
continuous patient monitoring in the face of increasing demands and limited medical resources. Through
their IoT-based framework, which incorporates various sensors to measure vital signs, they aim to help
the healthcare process, particularly in emergencies. By transmitting data to the cloud for analysis and
remote viewing, their system enables healthcare providers to stay informed and intervene promptly
when necessary.</p>
      <p>In [5], the authors explore the fusion of digital data processing with established medical techniques
to enhance patient care both at home and in clinical settings. They present various fall prediction and
detection methods, with a particular focus on vital signs monitoring using a triple-axis accelerometer
and parameters such as heart rate and temperature. The proposed system, VitaFALL, leverages the
Internet of Medical Things technology for timely fall detection and notification of caregivers and doctors
via GSM and GPRS modules. Despite observing some delays in detection, the system achieves promising
results, with an 85% accuracy rate and high specificity and sensitivity in detecting directional falls.</p>
      <p>Minglei et al. [6] present a real-time monitoring system designed to cater to the needs of the elderly,
individuals experiencing sub-optimal health, and those sufering from insomnia. This system enables
continuous monitoring of sleep quality and heart rate fluctuations, facilitating concentrated and efective
management of sleep states for various demographics. Utilizing vital signs monitoring devices, including
heart rate and breathing rate sensors, along with bed state detection, the system captures and transmits
data to a cloud platform server via WiFi. The server then conducts screening, analysis, and calculation
of vital signs data and dynamic changes, enabling individuals under monitoring to access their daily
vital signs and sleep state information through a mobile terminal. Managers can also monitor user data
and states in real-time, facilitating prompt detection of vital signs changes.</p>
      <p>In [7], the address the challenges faced by the healthcare sector in underfunded regions like
Bangladesh, aiming to bridge the gap by developing an afordable and accessible real-time health
monitoring system. Their solution utilizes Internet of Things (IoT) sensors connected to an Arduino
microprocessor to measure vital body signs, which are then transmitted to a smartphone via Bluetooth.
This data is further processed and displayed on an Android application interfaced with a web-based
platform, facilitating remote communication between patients and doctors in real-time. The system
underwent successful testing on multiple real-human subjects, demonstrating high accuracy in
monitoring vital signs such as body temperature, heart rate, ECG, SpO2 levels, blood pressure, and glucose
levels. Notably, the integration of IoT-based patient monitoring with telemedicine distinguishes this
system from existing solutions, ofering a comprehensive approach to remote healthcare management
and potentially improving global life expectancy.</p>
      <p>Our research uniquely focuses on developing a wearable patch-like sensor system for real-time
monitoring and alerting specifically tailored to post-attack scenarios, addressing the critical needs of
ifrst responders and victims in emergencies.</p>
    </sec>
    <sec id="sec-3">
      <title>3. System overview</title>
      <p>The system comprises a patch device that gathers data directly from an individual through multiple
sensors, conducting initial processing. The patch device collects ECG and PPG signals, and body
temperature. The primary computations carried out by the patch include calculating heart rate (HR) and
respiratory rate (RR) based on ECG signal processing and executing START triage (Fig. 2) while alerting
to any state changes. The algorithm used for HR and RR estimation is the Pan-Tompkins algorithm and
the details about it and the obtained results are presented in [8].</p>
      <p>Additionally, there is a tablet that receives data from the patch. The tablet estimates oxygen saturation
(SpO2) from PPG and blood pressure (BP) from ECG+PPG using trained AI models. The AI models for
SpO2 estimation are elaborated in [9], while for BP in [10].</p>
      <p>There is also a laser-induced graphene sensor (LIG) which is used also for HR and SpO2 estimation
as a backup and reference [11, 12].</p>
      <p>The simplified symbolic system design used in this project is shown in Figure 1. It represents the
basic concept of acquiring samples of diferent signals using various sensors and transferring them to
more computationally powerful devices for creating big data databases, processing the signals using
neural networks or artificial intelligence and more. To make the smart patch capable of doing this, it
needs the control logic which consists of low-power, inter-compatible hardware driven by well-designed
ifrmware (software that runs directly on the hardware).</p>
      <p>The patch assesses whether an individual’s vital signs (HR and RR) deviate from the anticipated
values, referencing the guidelines and values outlined in the triage process depicted in Figure 2. In the
event of a patient transitioning from a green to yellow or yellow to red status, the alarm is triggered,
accompanied by sound and the blinking of a light diode.</p>
      <p>The functions of the smart patch and their communication with the monitoring tablet were first
checked in laboratory conditions and the obtained values (Heart rate, Respiratory rate, Temperature,
Blood pressure, Oxygen saturation) were compared with data from reference devices. Based on the
tests, also the algorithms for BP and SpO2 estimation were improved.</p>
      <sec id="sec-3-1">
        <title>3.1. The Smart Patch Hardware</title>
        <p>The SmartPatch utilizes the MAX32630 ultra-low-power microcontroller, equipped with 2MB flash
memory and 512kB RAM. Its single Arm Cortex M4 core, featuring an FPU, operates at 96MHz.
Facilitating communication with the tablet, the PAN1780AT Bluetooth module is employed, supporting the
LR (long range) mode of BT v5.0 LE standard. This module is connected to the microcontroller through
an asynchronous serial interface (UART) with flow control and is managed through AT commands. It
includes MAX86150 ECG&amp;PPG sensor, laser-induced graphene sensor (LIG), STS35-DIS temperature
sensor, inertial measurement unit – BMI323, power management unit MAX77650, piezo speaker driver
PAM8904, and 4 high-intensity LEDs. The basic hardware block diagram, which also mimics the
ifrmware structure, is shown in Figure 3.</p>
        <p>Figure 4 illustrates the assembly of components on the Smart Patch. To construct an individual
patch, the control board must be flexibly linked with two holder boards. The battery and beeper will
be positioned above the control board and enclosed within a sealed, biocompatible, flexible packaging.
The complete patch will be fastened together using two snaps along with an adhesive strip containing
two ECG electrodes, an adhered LIG sensor, and an aperture for PPG and temperature sensors.</p>
        <p>For the sensor prototype of the smart patch, it was necessary to adapt existing firmware driver APIs
accessible online or to create new APIs if they were not available:
• API driver for temperature sensor STS35-DIS enables composing and sending configuration
bytes by calling representatively named functions. It also performs conversion from raw data to
temperature in °C;
• API driver for Inertial Measurement Unit BMI323 was developed. It is a reduced and completely
new API for configuring and data acquisition because the driver directly available from the
manufacturer did not fit our application, because incorporating all features of the device made
the driver slow;
• Simple API driver for piezo buzzer driver PAM8904 was created to simplify code writing;
• API driver for selected Power Management Unit MAX77650 was developed for comfort
programming of the IC. API is logically structured into separated blocks with various functions
corresponding to the IC’s internal hardware (Global Resources, Charger, LDO, Buck-Boost
Converter, and Current Sinks).</p>
      </sec>
      <sec id="sec-3-2">
        <title>3.2. Bluetooth Low Energy connection</title>
        <p>The system uses Bluetooth Low Energy for sending data to the tablet. To better understand the
complexity of Bluetooth low-energy communication, we will briefly explain the most important parts.
There are three communication methods, Advertising, Discovery, and Connection [13].</p>
        <p>Advertising is the public transfer of short data and server device information. Discovery is scanning
by client device which results in a list of nearby advertising server devices. Connection is a private
1-to-1 data exchange (with a selectable security level).</p>
        <p>In these communication methods, there are diferent BLE roles, Server and Client. The server device
(a smart patch in our case) can communicate using advertising (transmitting device information) and
connection. In connection, it can receive data from the client and notify (send data to) the client. The
client device (Android tablet in our case) can communicate using discovery (looking for advertising
devices) and connection. In connection, it can write data to the server and receive notifications from the
server if the client subscribes to them. The notifications are triggered by the value of the characteristic
in the server device. Subscribing to notifications is done by the client by writing to the corresponding
descriptor in the server device. The actual data transfer in connection happens in the value of the
characteristics. The BLE server device structure is quite complicated regarding extensive and constantly
updated Bluetooth Core Specifications. To keep this explanation short and to better understand these
previously mentioned terms, refer to the simple structure shown in Figure 5.</p>
        <p>The advertising data are public, thus anyone can receive them. However, the recipient still has to
know how to decode the bytes carrying the data. We transfer just the basic short data (ID, RR, HR . . . ) in
the advertising, but still to increase privacy, we use a Coded PHY (Coded Physical Layer) transfer layer
because devices are not usually discovered on this PHY. This PHY also brings another advantage of
longer communication distances. After connection, the transfer of all available data on the smart patch
starts only if the correct key or password request is sent from the client (tablet) to the server (patch).</p>
        <p>The advertised data by the patch at this point are patch ID, status bit field representing the state of
the patch sensors and algorithms, heart rate, respiratory rate, body surface temperature, patch battery
voltage, and version of the patch firmware.</p>
      </sec>
      <sec id="sec-3-3">
        <title>3.3. Software Application</title>
        <p>The Android application is developed in Android Studio IDE [14] using Kotlin language [15]. The GUI
uses Jetpack Compose [16]. The Bluetooth Low Energy interface is from Standard Android Libraries to
have access to the newest implemented features. For compatibility with Python modules for the SpO2
and blood pressure estimation, ChaquoPY [17] is used. It is necessary for the simplified development of
artificial intelligence predictive neural network models in Python.</p>
        <p>First, a short brief of the main application capabilities:
• one Android device can show multiple advertising patches;
• one Android device can connect to multiple patches;
• any Android device can connect to any patch;
• data can be stored in a file for later examination;
• values on the screen are updated each second;
• battery and Bluetooth signal quality information;
• measured signals (ECG, PPG) can be displayed in a selectable timebase;
• AI predictive models are computed in parallel threads;
• received measured data are converted and processed in parallel threads.</p>
        <p>The most complex part of the application is the Bluetooth Low Energy backend. In discovery mode,
it scans for advertising devices. When the BLE server device is found, it checks if it has the correct
device name prefix if its extended advertising includes the correct service data structure, and if the
value of this structure has a minimal required length. If all these conditions are satisfied, the BLE device
is recognized and it is added on the screen or if it is already there, values sent by it are updated. In
connection mode, the low-level hardware connection establishment happens "in the background".</p>
        <p>Then, there are actions that the code must perform using the GATT (Generic Attribute Profile)
interface. If a connection is established, the application discovers services in the BLE server. If there is
the correct service in the "on discover services" callback, it discovers characteristics. If there are correct
characteristics in service, it reads and updates the value of the CCCD descriptor (Client Characteristic
Configuration Descriptor) of one of the characteristics. If the descriptor is successfully written in the
"on descriptor write" callback, it writes a password request to the other characteristics, indicating the
subscription for notifications. Then, the data will be received from the patch in the "on characteristic
changed" callback.</p>
        <sec id="sec-3-3-1">
          <title>3.3.1. Application beta testing</title>
          <p>The application uses Bluetooth LE Long Range Coded Physical Layer for both discovery and connection
mode. When the application starts, it checks if it has all the necessary permissions. If they are not
allowed, the application asks the user for their approval. If the Bluetooth feature of the device is disabled,
the application prompts the user to allow it to turn the Bluetooth on.</p>
          <p>The application can run with both, the portrait and landscape orientation of the screen. The user
interface has control buttons at the top of the screen. There are buttons for controlling the Bluetooth
LE discovery (Start scan – Stop scan). If there are more available smart patches, they will appear on the
screen. Then there is a button to enable or disable storing of received raw data after connection.</p>
          <p>In the top right part of the screen, there are buttons for selecting the plot time base of displayed ECG
and PPG signals on connected patch panels. Most to the right there are two switches: the first one
changes the behavior of the patch panels. When the switch is checked, only the ID field in the patch
panels acts as a button to change the state of the connection (connect to the smart patch or disconnect
from the patch). When the switch is unchecked, the whole patch panel is clickable. Thus, clicking
anywhere inside the patch panel will cause the connection change. In some cases, the medical personnel
do not have enough time to hit precisely one button, so they can touch the patch panel anywhere. On
the other hand, when the user wants to point out some features of the displayed signals or values, he
might accidentally touch the screen and cause unwanted patch disconnection. This is the reason for
this control element, so the users can adapt the control behavior based on the situation. The second
switch turns on/of the display of a short patch ID on the top of the patch panel.</p>
        </sec>
        <sec id="sec-3-3-2">
          <title>3.3.2. Patch panels</title>
          <p>When the smart patch is discovered, its panel displays just the values that are computed or measured on
the patch. Currently, these are the heart rate (HR), respiratory rate (RR), and body surface temperature
(BT). The blood pressure (BP) and SpO2 are computed in the Android device (tablet) using artificial
intelligence predictive neural network models, so they are displayed only after the Android device
connects to the patch (and reads the raw data from the patch). It is the same with plots of ECG and PPG
signals: they are displayed only during connection because the signal data are streamed in real-time
only when the patch is connected. The smart patch panel background is white when it is discovered,
not connected, and the HR and RR are in the "yellow" patient limits of the START triage procedure.
The vital signs are black in default. When HR and/or RR are/is out of bounds of these limits, the panel
background turns red to indicate that the patient needs medical attention. The out-of-bounds vital sign
value will change its color to white (on a red panel background). When the patch is connected, its panel
background changes to green to indicate the active connection. During connection, the out-of-bounds
values will be red (on the green tab background). Some screenshots from the application documenting
the above-mentioned features and functions are in Figures 6 and 7.</p>
          <p>The development of mathematical models for health status was integrated into the smart patch
processor using the C programming language, specifically through the START procedure. This integration
facilitates the generation of alerts when the health status transitions from the Yellow label to the Red
label is identified.</p>
          <p>Additionally, the most efective artificial intelligence (AI) models were selected and incorporated into
the Android application. This enables the AI predictive neural network (NN) models for blood pressure
(BP) and SpO2 estimation to undergo computation in parallel threads. This parallel processing capability
enhances real-time processing power, enabling the extraction of desired features in real-time. The most
important task of the Smart patch is to perform part of the triage procedure continuously in real-time.
If the respiratory rate is in bounds from 10 to 30 breaths per minute (including) and the heat rate is in
bounds from 40 to 120 beats per minute including, in other words:  ∈ &lt;10; 30&gt; ∧  ∈ &lt;40; 120&gt;,
the patient status is Yellow. If this condition is not satisfied, hence HR or RR is out of these bounds,
the patient status is red. In the tablet application, there is a programmable hysteresis to prevent the
possible rounding error to cause false fast switching between states. The hysteresis of one count (±1)
has been used. If there is a change in patient triage status, the appropriate LED is lighted up (red or
yellow) and the other one is turned of (yellow or red) and the corresponding sequence of tones starts
to play for the required amount of time.</p>
        </sec>
      </sec>
    </sec>
    <sec id="sec-4">
      <title>4. Conclusion</title>
      <p>In this paper, we have presented a comprehensive approach to the development of a Smart System
for Real-time Monitoring of Vital Parameters. Our research addresses the critical need for continuous
monitoring of vital signs in both military and civilian contexts, particularly during high-stress situations
such as terrorist attacks, IED incidents, or rescue operations.</p>
      <p>We proposed the design and development of a prototype patch-like device capable of gathering
and analyzing vital parameters including respiration rate, heart rate, blood oxygen saturation, blood
pressure, and body temperature in real-time. Through the integration of Artificial Intelligence (AI)
models, our system enables the prediction of blood pressure and oxygen saturation and issues alerts in
the event of any health status changes, thus assisting medical personnel in making timely decisions.</p>
      <p>The system incorporates advanced features such as Bluetooth Low Energy connectivity, real-time
data processing, and graphical user interfaces for visualization of vital signs. Furthermore, we have
demonstrated the eficacy of our system through laboratory testing and validation, comparing the
measured data with reference devices and refining algorithms for improved accuracy.</p>
      <p>Our research lays the groundwork for further innovation in this field, with ongoing eforts focused
on refining the hardware, optimizing software algorithms, and conducting real-world trials to validate
the efectiveness of our system in diverse operational environments.</p>
      <p>However, our study has several limitations. The prototype has been primarily tested in controlled
laboratory environments, and its performance in real-world scenarios remains to be thoroughly
evaluated. Additionally, the system’s reliance on Bluetooth connectivity may pose challenges in areas
with significant interference or limited connectivity. The AI models require further optimization and
extensive training with larger datasets to enhance their predictive accuracy and reliability.</p>
      <p>Future research will focus on refining the hardware, optimizing software algorithms, and conducting
real-world trials to validate the system’s efectiveness in diverse operational environments. This will
include integrating additional sensors, developing more sophisticated AI algorithms, and implementing
robust security measures to ensure data privacy and integrity.</p>
    </sec>
    <sec id="sec-5">
      <title>Acknowledgments</title>
      <p>This work was supported in part by the NATO Science for Peace and Security Program under project
SP4LIFE, number G5825. This work was supported in part by the Faculty of Computer Science and
Engineering in Skopje, North Macedonia under project BIOX.
[2] M. H. Yousefi, B. Behnam, S. Farahani, An auxiliary framework to facilitate earthquake search and
rescue operations in urban regions, Natural Hazards (2024) 1–25.
[3] J. M. Franc, S. W. Kirkland, U. D. Wisnesky, S. Campbell, B. H. Rowe, Metastart: a systematic
review and meta-analysis of the diagnostic accuracy of the simple triage and rapid treatment
(start) algorithm for disaster triage, Prehospital and disaster medicine 37 (2022) 106–116.
[4] V. Yeri, D. Shubhangi, Iot based real time health monitoring, in: 2020 Second international
conference on inventive research in computing applications (ICIRCA), IEEE, 2020, pp. 980–984.
[5] W. Patel, C. Patel, B. Ramani, S. Bhaskar, M. Patel, Vitafall: Nxtgeuh system for well-being
monitoring with fall recognition and real-time vital sign monitoring, Int J Recent Technol Eng
(IJRTE) 8 (2020) 5139–5145.
[6] M. Shu, M. Tang, M. Yang, N. Wei, The vital signs real-time monitoring system based on internet
of things, in: 2017 4th International Conference on Information Science and Control Engineering
(ICISCE), IEEE, 2017, pp. 747–751.
[7] A. Ahmed, M. M. Khan, P. Singh, R. S. Batth, M. Masud, Iot-based real-time patients vital
physiological parameters monitoring system using smart wearable sensors, Neural Computing
and Applications 34 (2022) 19397–19673.
[8] D. Gogola, N. Ackovska, R. Bagín, B. Koteska, A. M. Bogdanova, M. Kostoska, F. Lehocki, M. Tysler,
Integrated smart patch for heart rate and respiratory rate monitoring, in: 2023 14th International
Conference on Measurement, IEEE, 2023, pp. 63–66.
[9] B. Koteska, A. M. Bodanova, H. Mitrova, M. Sidorenko, F. Lehocki, A deep learning approach
to estimate spo2 from ppg signals, in: Proceedings of the 9th International Conference on
Bioinformatics Research and Applications, 2022, pp. 142–148.
[10] I. Kuzmanov, A. M. Bogdanova, M. Kostoska, N. Ackovska, Fast cufless blood pressure classification
with ecg and ppg signals using cnn-lstm models in emergency medicine, in: 2022 45th Jubilee
International Convention on Information, Communication and Electronic Technology (MIPRO),
IEEE, 2022, pp. 362–367.
[11] T. Vićentić, M. Rašljić Rafajilović, S. D. Ilić, B. Koteska, A. Madevska Bogdanova, I. A. Pašti,
F. Lehocki, M. Spasenović, Laser-induced graphene for heartbeat monitoring with heartpy analysis,
Sensors 22 (2022) 6326.
[12] A. Madevska Bogdanova, B. Koteska, T. Vićentić, S. D Ilić, M. Tomić, M. Spasenović, et al., Blood
oxygen saturation estimation with laser-induced graphene respiration sensor, Journal of Sensors
2024 (2024).
[13] C. Liu, Y. Zhang, H. Zhou, A comprehensive study of bluetooth low energy, in: Journal of Physics:</p>
      <p>Conference Series, volume 2093, IOP Publishing, 2021, p. 012021.
[14] Google, Android studio, https://developer.android.com/studio, 2013. Accessed: 2024-05-20.
[15] JetBrains, Kotlin, https://kotlinlang.org, 2011. Accessed: 2024-05-20.
[16] Google, Jetpack compose, https://developer.android.com/jetpack/compose, 2020. Accessed:
202405-20.
[17] Chaquo, Chaquopy, https://chaquo.com/chaquopy/, 2018. Accessed: 2024-05-20.</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <ref id="ref1">
        <mixed-citation>
          [1]
          <string-name>
            <given-names>S.</given-names>
            <surname>Mondal</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Zehra</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A.</given-names>
            <surname>Choudhury</surname>
          </string-name>
          ,
          <string-name>
            <given-names>P. K.</given-names>
            <surname>Iyer</surname>
          </string-name>
          ,
          <article-title>Wearable sensing devices for point of care diagnostics</article-title>
          ,
          <source>ACS Applied Bio Materials</source>
          <volume>4</volume>
          (
          <year>2020</year>
          )
          <fpage>47</fpage>
          -
          <lpage>70</lpage>
          .
        </mixed-citation>
      </ref>
    </ref-list>
  </back>