        <article-title>Comparative Plausibility Evaluation of Heatmaps for Vision Transformers in Digital Mammography</article-title>
      </title-group>
      <contrib-group>
        <contrib contrib-type="author">
          <string-name>Ramsha Aasim</string-name>
          <email>raasim@umd.edu</email>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Ghada Zamzmi</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Jana G. Delfino</string-name>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Joseph Jaja</string-name>
          <xref ref-type="aff" rid="aff1">1</xref>
        </contrib>
        <contrib contrib-type="author">
          <string-name>Miguel A. Lago</string-name>
          <email>miguel.lago@fda.hhs.gov</email>
          <xref ref-type="aff" rid="aff0">0</xref>
        </contrib>
        <aff id="aff0">
          <label>0</label>
          <institution>Ofice of Science and Engineering Laboratories, Center for Devices and Radiological Health, U.S. Food and Drug Administration</institution>
          ,
          <addr-line>Silver Spring, MD</addr-line>
          ,
          <country country="US">USA</country>
        </aff>
        <aff id="aff1">
          <label>1</label>
          <institution>University of Maryland's Institute of Health Computing</institution>
          ,
          <addr-line>North Bethesda, MD</addr-line>
          ,
          <country country="US">USA</country>
        </aff>
      </contrib-group>
      <abstract>
        <p>To efectively integrate Artificial Intelligence (AI) into healthcare workflows, robust AI explainability (XAI) techniques are needed to build clinician trust and understanding of AI-driven predictions. With numerous XAI methods available, objectively determining the most suitable method for a given task - beyond subjective human assessment -remains a challenge. This study tackles this challenge by introducing an evaluation framework that enables explainability assessments to foster trust and accelerate AI adoption in clinical settings. We demonstrate the plausibility criterion - defined as the degree to which AI explanations align with established ground truth for a given task -as a means to quantitatively assess XAI. To illustrate our evaluation criteria in medical imaging, we compared five post hoc heatmap methods applied to a Vision Transformer (ViT) trained for lesion detection in digital mammography. For each technique, we generate and compare heatmaps, and then examine how varying the mammogram's radiation dose afect their plausibility. Our analysis shows how this framework quantitatively measures each heatmap method's ability to highlight diagnostically relevant regions, providing an objective way to evaluate AI explanation quality. We conclude by emphasizing that, while plausibility is a key metric, it should be considered alongside other relevant criteria. This work is part of our broader eforts to develop a comprehensive framework to quantitatively assess the quality of AI explanations in the regulatory evaluation of medical imaging devices.</p>
      </abstract>
      <kwd-group>
        <kwd>eol&gt;AI Explainability (XAI)</kwd>
        <kwd>Regulatory Evaluation</kwd>
        <kwd>Vision Transformers (ViTs)</kwd>
        <kwd>Heatmaps</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec id="sec-1">
      <title>1. Introduction</title>
      <p>
        Vision Transformers (ViTs) [
        <xref ref-type="bibr" rid="ref1">1</xref>
        ] have rapidly gained prominence in computer vision as they ofer a
compelling alternative to Convolutional Neural Networks (CNNs) for tasks such as image classification,
segmentation, and object detection. ViTs primarily utilize the self-attention mechanism [
        <xref ref-type="bibr" rid="ref2">2</xref>
        ], in order to
capture global dependencies, ofering minimal inductive bias. In medical imaging, ViTs have proved
[
        <xref ref-type="bibr" rid="ref3">3</xref>
        ] to perform well for applications such as segmentation [4], detection [5], classification [ 6], and
reconstruction [7].
      </p>
      <p>Despite their widespread popularity, transformer models have a complex architecture, making it
dificult to understand and explain their predictions. Explainable AI (XAI) aims to tackle this issue by
ofering techniques that provide insights into the model’s reasoning [ 8, 9, 10]. However, the field still
lacks a comprehensive and standardized framework for evaluating the efectiveness of these explanation
techniques [11, 12].</p>
      <p>This study proposes an evaluation framework specifically designed to assess the efectiveness of
heatmap-based explanations for ViTs. In particular, we focus on heatmap-based explanations generated
by attribution methods, which assign a relevance score to each input feature (image pixels, in our case)
based on its contribution to the final prediction; a higher positive value indicates a greater impact
on the model’s prediction. While each explainability method is designed to generate ’localized’ and
’relevant’ explanations, the resulting heatmaps often vary significantly. This inconsistency, along
with the lack of objective evaluation techniques for these generated explanations, makes it dificult
to confidently rely on and trust these explanations in practice. To address this challenge, our study
explores “plausibility”—defined as the alignment of a generated heatmap with an established ground
truth—as a criterion to assess the quality and localization of heatmaps. To demonstrate this approach,
we evaluate and compare the plausibility of heatmaps generated by five AI explainability methods
applied to a ViT trained for lesion detection in digital mammography data.</p>
    </sec>
    <sec id="sec-2">
      <title>2. Related Work</title>
      <p>One of the most widely used explainability methods is heatmaps, which are designed to visualize the
regions of input data that contribute most significantly to a model’s prediction. Heatmaps have become
an essential tool for explaining CNNs and highlighting discriminative regions in the input. CAM
based techniques are among the most prevalent explainability techniques for heatmap generation for
CNNs [8, 13, 14].These explainability have since been extented to Vision Transformers (ViTs) [15]. The
main component of a ViT is the multi-head self-attention mechanism to model relationships between
patch embeddings. To that end, several attention-based methods have also been introduced, including
raw attention visualizations, the attention-rollout technique [9], which aggregates information from
diferent attention heads across transformer blocks, and relevance-based explainability techniques
[10, 16] that incorporate the relevance of attention heads in their explanations. Perturbation-based
methods, such as ViT-CX [17], have also been utilized for generating explanations.</p>
      <p>While the field of XAI is rapidly advancing, the research around the quantitative evaluation of
these explanations remains less explored. This gap is particularly evident in medical imaging, where
explanations often lack thorough evaluation [18]. For instance, Lysdahlgaard [19] compares various AI
models on wrist and hand X-ray data based on the explanations generated by GradCAM. However, in
the absence of standardized evaluation protocols, there is no way to tell if the GradCAM is actually
generating the “right” explanations and if such explanations can be trusted. Some research studies
suggest various techniques to evaluate explanations, based on sensitivity, complexity and faithfulness
of explanations to the model. [20, 12, 21]. However, the field lacks a consensus on how to quantitatively
assess AI explainability methods, often resorting to illustrative but potentially subjective examples
[22]. Our work aims to contribute to establishing a standardized framework for evaluating heatmap
explanations. We propose plausibility as a key evaluation criterion and explore diferent metrics for its
quantification. Although the concept of evaluating heatmaps based on their ability to highlight relevant
regions has been touched upon in general XAI research [23, 24], our work specifically investigates its
application within the domain of medical imaging.</p>
    </sec>
    <sec id="sec-3">
      <title>3. Methods</title>
      <p>In this section, we outline the methodology used to quantitatively evaluate the plausibility of
heatmapbased explainability methods applied to ViTs in the context of medical imaging.</p>
      <sec id="sec-3-1">
        <title>3.1. Heatmap Explanations for ViT</title>
        <p>We applied three CAM-based methods and two attention-based methods for explainability. For the
classdiscriminative explainability methods, the explanations are generated based on the model’s predicted
class.</p>
        <sec id="sec-3-1-1">
          <title>3.1.1. CAM based Methods</title>
          <p>Class Activation Map (CAM) explainability techniques were initially designed to explain CNN
predictions. The core idea is that the final convolutional layer’s feature maps retain information about
image regions crucial for classifying the target category. A heatmap is then generated by taking a
weighted combination of these feature maps and projecting it back onto the input image to highlight
important regions. Diferent CAM methods vary in how they calculate the weights for these feature
maps. For example, GradCAM [8], a foundational and widely used approach, calculates the gradient of
each feature map with respect to the target class and then the weights for each feature map is calculated
to be the average of the gradients of that feature map. GradCAM Element-wise [25], is a variation of
GradCAM, but in contrast, weighs individual feature map values with their corresponding gradient
values before combining them. EigenGradCAM, an extension of EigenCAM [13], is another method that
leverages the eigenvalues of the element-wise gradient weighted feature maps to generate heatmaps.</p>
          <p>The utility of CAM methods has also been extended to explain ViT models as illustrated in Figure
1. In contrast to CNNs, ViTs process images as a sequence of fixed size, non-overlapping patches,
which are embedded and subsequently transformed through multiple self-attention-based transformer
blocks. Each patch is represented by a vector embedding of dimension D, a dimensionality that is
maintained throughout the transformer blocks. To adapt CAM techniques for ViTs, we extract patch
representations from an intermediate transformer layer, typically the second to last due to the lack
of gradient flow in the final layer. These patch vectors are then rearranged to reflect their original
spatial arrangement in the image, with the embedding dimension D forming the third dimension. This
reshaped representation is then treated as a set of D feature maps, enabling the application of standard
CAM methods by determining a suitable weighting scheme for their combination (Figure 1). For our
experiments, we apply the GradCAM, EigenGradCAM and GradCAM Element-wise techniques on
Vision Transformers.</p>
        </sec>
        <sec id="sec-3-1-2">
          <title>3.1.2. Attention based Methods</title>
          <p>For the attention-based explanations, we use the attention rollout method and the relevance-based
explainability method.</p>
          <p>Attention Rollout [9] is a label-agnostic technique that consolidates the flow of information within
the Vision Transformer’s attention mechanism. In each transformer block, multiple attention heads
assign attention scores based on the importance of individual image patches for propagating relevant
information to the next layer, creating a set of attention matrices that focus on diferent aspects of the
input. The rollout method then aggregates these attention matrices—using operations such as averaging,
minimum, or maximum across the attention heads—and incorporates the identity matrix (I) to account
for residual connections. The resulting attention matrix (A) for each transformer block is recursively
multiplied across all layers to produce a final matrix, which is used for generating the heatmap.</p>
          <p>The Transformer Multi-Modal Explainability method [16], extends the Layerwise Relevance
Propagation-Based Method, Beyond Attention [10]. It utilizes the relevance matrices for each attention
head instead of raw attention scores. These matrices are weighted by their corresponding gradients
(discarding negative values), and the resulting positive relevance is propagated across transformer
layers, similar to Attention Rollout.</p>
        </sec>
      </sec>
      <sec id="sec-3-2">
        <title>3.2. Heatmap Plausibility Evaluation</title>
        <p>The plausibility of an AI model’s explanation refers to how well it aligns with established ground truth,
leveraging relevant background knowledge or general consensus. This is also known as “coherence"
[22] or “localization" [23]. In medical imaging, a plausible explanation, such as a heatmap, should
highlight clinically significant regions that correspond to a clinician’s area of focus for a given task. For
instance, in lesion detection, a plausible heatmap would emphasize the lesion’s location.</p>
        <p>To quantify plausibility, an ideal heatmap should assign high attribution values to pixels within
the ground truth region and low values to pixels outside this area. We use three metrics to evaluate
plausibility in our analysis.</p>
        <sec id="sec-3-2-1">
          <title>3.2.1. Intersection over Union</title>
          <p>The Intersection over Union (IoU) [26] metric measures the spatial overlap between the ground truth
lesion annotation and the predicted high-intensity region derived from the heatmap. This predicted
region is determined by identifying the pixel with the highest attribution value and creating a bounding
box of the same dimensions as the ground truth, with the high intensity point as the center. The IoU
ranges from 0 (no overlap) to 1 (perfect overlap), with higher values indicating better plausibility.</p>
        </sec>
        <sec id="sec-3-2-2">
          <title>3.2.2. Relevance Mass Accuracy</title>
          <p>Relevance Mass Accuracy quantifies the proportion of total attribution mass concentrated within the
ground truth region. Given a heatmap with at least one positive attribution value, Arras et al. [24]
defines it as the ratio of the sum of the positive attribution values  of the pixels that lie inside the
ground truth (GT) region to the sum of attributions  of the total pixels N in the heatmap. It is
mathematically given by:
 = | ∩  |</p>
          <p>| |
  = ∑∑︀︀∈ 
=1</p>
          <p>The Relevance Mass Accuracy will have a value of 1 if all positive attributions of the heatmap lie
inside the identified ground truth region.</p>
        </sec>
        <sec id="sec-3-2-3">
          <title>3.2.3. Relevance Rank Accuracy</title>
          <p>The relevance rank accuracy measures the number of pixels in the ground truth region that have high
positive attribution values. To calculate the high attribution values, we maintain the top 10% of the
attribution values in the heatmap and set the rest to 0. Let  be the set of pixels that contain the top
k=10% of the attributions and GT is the set of all the pixels that lie inside the ground truth, the rank
accuracy is given by:
(1)
(2)</p>
        </sec>
      </sec>
      <sec id="sec-3-3">
        <title>3.3. Experimental Setup</title>
        <p>3.3.1. Dataset
We used M-SYNTH dataset [27] for training the lesion detection model. M-SYNTH, developed at the
U.S. Food and Drug Administration (FDA), is a synthetic dataset of 45,000 digital mammography images
constructed using stochastic knowledge-based models to simulate realistic variations based on the
breast fibroglandular density, lesion size and density, and radiation dose. M-SYNTH mammograms
include both lesion-containing and lesion-free images. One of the key advantages of M-SYNTH is its
precise annotation of lesion locations, provided in the form of pixel-level masks that accurately delineate
the lesion regions. In this study, we use bounding boxes derived from these masks as the ground truth.
This reliable ground truth enables a meaningful evaluation of how efectively explainability methods
can highlight the relevant areas within the image.</p>
        <p>The data used was a subset of the M-SYNTH dataset consisting of 900 images with fatty breast density,
100% recommended radiation dose, and containing lesions with diameters of 5.0, 7.0, and 9.0 mm (equally
distributed) and a lesion density of 1.06. This dataset maintained a balanced 50/50 distribution of images
with and without lesions. We divided this dataset into 70% training, 15% tuning, and 15% testing.</p>
        <p>To evaluate the heatmaps for plausibility, we selected a random sample of 150 lesion-present
mammogram images from the subset of M-SYNTH data with equal representations of lesions with sizes 5.0,
7.0, and 9.0.</p>
        <sec id="sec-3-3-1">
          <title>3.3.2. Model Training and Evaluation</title>
          <p>
            We employed a vanilla ViT [
            <xref ref-type="bibr" rid="ref1">1</xref>
            ] as a binary classifier to detect the presence or absence of lesions in
mammograms. The lesion detection is labeled with "WITHLESION" or "NOLESION". We fine-tuned the
"vit-base-patch16-224" model, pre-trained on ImageNet [28] and available via Hugging Face, using an
input image of 224 × 224 and a patch size of 16 × 16. Prior to training and inference, the input images
were preprocessed by normalizing pixel values with a mean and standard deviation of 0.5 and resized
to 224 × 224 pixels. We trained the model for 7 epochs with a batch size of 12 and a learning rate of
6 × 10− 6. The model achieved a test accuracy of 99.3% and an AUC score of 0.99, which is important
for ensuring that the visual explanations generated by heatmaps are meaningful.
          </p>
        </sec>
        <sec id="sec-3-3-2">
          <title>3.3.3. Target Layer Selection for CAM methods</title>
          <p>CNNs process images through convolutional blocks, inherently preserving spatial information. Thus,
leveraging the last convolutional layer, which carries significant discriminatory information and is
subsequently used for classification via fully connected layers (Multi-Layer Perceptron head, MLP), is
a logical choice for CAM based techniques. In contrast, the self-attention layers in ViTs inject global
information into each patch as it propagates through the network. As a result, the patch representations
evolve to not only contain information about their own spatial location but also about other patches in
the image. Consequently, towards the end of the network, these patch representations tend to become
more abstract [9, 29], potentially rendering CAM methods that rely on the final layers less interpretable.
Therefore, selecting the appropriate target layer for the generation of ViT heatmaps requires careful
consideration. We explore various ViT transformer block outputs as the target layers, across the three
CAM methods. We present our results in Figure 2, observing that earlier layers tend to emphasize the
lesion, while later layers result in heatmaps that are more dispersed and less focused.</p>
          <p>In order to determine the optimal target layer for our experiments, we evaluate the average plausibility
score achieved by the three CAM methods across our balanced sample set. The results for this experiment
are shown in Figure 3. The target layer that yielded the highest average score across all CAM methods
i.e., output layer for the sixth transformer block, was selected for subsequent analysis for the CAM
based methods.</p>
        </sec>
      </sec>
    </sec>
    <sec id="sec-4">
      <title>4. Results</title>
      <p>Figure 4 presents a comparative analysis of the average plausibility scores for the explainability
methods, GradCAM, EigenGradCAM, GradCAM Element-wise, Attention Rollout, and the Transformer
Multi-Modal Explainability method, evaluated on our sample set. Notably, GradCAM yields very
low plausibility scores for all the metrics, close to zero, indicating that its generated heatmaps are
not plausible, and are broadly dispersed rather than localized. The other methods exhibit a range of
plausibility scores, with EigenGradCAM and GradCAM Element-wise achieving the highest values. An
interesting observation is that while EigenGradCAM and GradCAM Element-wise exhibit similar values
for IoU and relevance rank accuracy—suggesting comparable performance in identifying the ground
truth region—EigenGradCAM achieves a higher relevance mass accuracy. This indicates that although
both methods highlight the correct region, GradCAM Element-wise also attributes high relevance
scores to surrounding breast tissue, leading to less precise localization. A similar pattern is observed
when comparing the attention rollout method with the Transformer MM-Explainability method. While
attention rollout yields higher IoU and relevance rank accuracy, the MM-Explainability method achieves
better relevance mass accuracy, suggesting more focused and localized heatmaps. Such analysis can
help identify the strengths and weaknesses of various explainability methods.</p>
    </sec>
    <sec id="sec-5">
      <title>5. Discussion and Conclusion</title>
      <p>This study introduces plausibility as a key evaluation criterion for explainability methods in medical
imaging. We quantitatively assessed the plausibility of explanations generated by a ViT model trained
for lesion detection in mammograms, comparing five post hoc heatmap techniques. Our results highlight
the significance of plausibility, measured using three metrics: IoU with ground truth, Relevance Mass
Accuracy, and Relevance Rank Accuracy. These metrics enable objective comparison of heatmaps and
IoU
MA
RA
provide insight into how well AI-generated explanations align with clinically meaningful regions, which
is an important consideration for building trustworthy and interpretable medical AI systems.</p>
      <p>A key limitation of plausibility evaluation is its reliance on a well-defined ground truth. In our
study, the use of a synthetic dataset allowed us to access precise ground truth lesion annotations, only
available for lesion-present trials. However, in some real-world clinical tasks, ground truth is often
subjective and based on expert interpretation. This makes plausibility harder to define and measure.
Further, plausibility is less applicable to global tasks such as breast density assessment, where no specific
localized ground truth exists. Another challenge arises when models are trained on poor-quality data
and fail to learn meaningful patterns. In such cases, an explanation that does not align with the ground
truth might still be valuable, as it could reveal the model’s limitations. Therefore, while plausibility
is a useful evaluation criterion, it should be interpreted alongside other criteria to provide a more
comprehensive assessment of explanation quality. Examples of other criteria that might need to be
considered when assessing explanation quality include consistency, fidelity, and usefulness. In future
work, we aim to incorporate a broader range of evaluation criteria for ViT explainability to provide a
more complete and nuanced evaluation of the generated heatmaps.</p>
      <p>This work underscores the importance of quantitative evaluation methods in selecting appropriate
XAI methods for medical applications and contributes to the broader goal of developing a comprehensive
framework for assessing the quality of explanations provided in medical imaging tasks.</p>
    </sec>
    <sec id="sec-6">
      <title>Declaration on Generative AI</title>
      <p>For the preparation of this work, the authors have not employed any Generative AI tools.
[4] M. Heidari, A. Kazerouni, M. Soltany, R. Azad, E. K. Aghdam, J. Cohen-Adad, D. Merhof, Hiformer:
Hierarchical multi-scale representations using transformers for medical image segmentation, in:
Proceedings of the IEEE/CVF winter conference on applications of computer vision, 2023, pp.
6202–6212.
[5] Y. Wu, Q. Kong, L. Zhang, A. Castiglione, M. Nappi, S. Wan, Cdt-cad: Context-aware deformable
transformers for end-to-end chest abnormality detection on x-ray images, IEEE/ACM Transactions
on Computational Biology and Bioinformatics (2023).
[6] X. Huo, G. Sun, S. Tian, Y. Wang, L. Yu, J. Long, W. Zhang, A. Li, Hifuse: Hierarchical multi-scale
feature fusion network for medical image classification, Biomedical Signal Processing and Control
87 (2024) 105534.
[7] A. Luthra, H. Sulakhe, T. Mittal, A. Iyer, S. Yadav, Eformer: Edge enhancement based transformer
for medical image denoising, arXiv preprint arXiv:2109.08044 (2021).
[8] R. R. Selvaraju, M. Cogswell, A. Das, R. Vedantam, D. Parikh, D. Batra, Grad-cam: Visual
explanations from deep networks via gradient-based localization, in: Proceedings of the IEEE international
conference on computer vision, 2017, pp. 618–626.
[9] S. Abnar, W. H. Zuidema, Quantifying attention flow in transformers, CoRR abs/2005.00928 (2020).</p>
      <p>URL: https://arxiv.org/abs/2005.00928. arXiv:2005.00928.
[10] H. Chefer, S. Gur, L. Wolf, Transformer interpretability beyond attention visualization, 2021. URL:
https://arxiv.org/abs/2012.09838. arXiv:2012.09838.
[11] D. Gunning, E. Vorm, J. Y. Wang, M. Turek, Darpa’s explainable ai (xai)
program: A retrospective, Applied AI Letters 2 (2021) e61. URL: https://onlinelibrary.
wiley.com/doi/abs/10.1002/ail2.61. doi:https://doi.org/10.1002/ail2.61.
arXiv:https://onlinelibrary.wiley.com/doi/pdf/10.1002/ail2.61.
[12] C.-K. Yeh, C.-Y. Hsieh, A. S. Suggala, D. I. Inouye, P. Ravikumar, On the (in)fidelity and sensitivity
for explanations, 2019. URL: https://arxiv.org/abs/1901.09392. arXiv:1901.09392.
[13] M. B. Muhammad, M. Yeasin, Eigen-cam: Class activation map using principal components, in:
2020 International Joint Conference on Neural Networks (IJCNN), IEEE, 2020, p. 1–7. URL: http:
//dx.doi.org/10.1109/IJCNN48605.2020.9206626. doi:10.1109/ijcnn48605.2020.9206626.
[14] H. Wang, Z. Wang, M. Du, F. Yang, Z. Zhang, S. Ding, P. Mardziel, X. Hu, Score-cam: Score-weighted
visual explanations for convolutional neural networks, 2020. URL: https://arxiv.org/abs/1910.01279.
arXiv:1910.01279.
[15] S. Stassin, V. Corduant, S. A. Mahmoudi, X. Siebert, Explainability and evaluation of vision
transformers: An in-depth experimental study, Electronics 13 (2024). URL: https://www.mdpi.
com/2079-9292/13/1/175. doi:10.3390/electronics13010175.
[16] H. Chefer, S. Gur, L. Wolf, Generic attention-model explainability for interpreting bi-modal and
encoder-decoder transformers, 2021. URL: https://arxiv.org/abs/2103.15679. arXiv:2103.15679.
[17] W. Xie, X.-H. Li, C. C. Cao, N. L. Zhang, Vit-cx: Causal explanation of vision transformers, 2023.</p>
      <p>URL: https://arxiv.org/abs/2211.03064. arXiv:2211.03064.
[18] P. Komorowski, H. Baniecki, P. Biecek, Towards evaluating explanations of vision transformers
for medical imaging, in: 2023 IEEE/CVF Conference on Computer Vision and Pattern Recognition
Workshops (CVPRW), IEEE, 2023, p. 3726–3732. URL: http://dx.doi.org/10.1109/CVPRW59228.2023.
00383. doi:10.1109/cvprw59228.2023.00383.
[19] S. Lysdahlgaard, Utilizing heat maps as explainable artificial intelligence for detecting abnormalities
on wrist and elbow radiographs, Radiography 29 (2023) 1132–1138.
[20] U. Bhatt, A. Weller, J. M. F. Moura, Evaluating and aggregating feature-based model explanations,
2020. URL: https://arxiv.org/abs/2005.00631. arXiv:2005.00631.
[21] D. Alvarez-Melis, T. S. Jaakkola, On the robustness of interpretability methods, 2018. URL: https:
//arxiv.org/abs/1806.08049. arXiv:1806.08049.
[22] M. Nauta, J. Trienes, S. Pathak, E. Nguyen, M. Peters, Y. Schmitt, J. Schlötterer, M. van Keulen,
C. Seifert, From anecdotal evidence to quantitative evaluation methods: A systematic review on
evaluating explainable ai, ACM Computing Surveys 55 (2023) 1–42. URL: http://dx.doi.org/10.
1145/3583558. doi:10.1145/3583558.
[23] A. Hedström, L. Weber, D. Bareeva, D. Krakowczyk, F. Motzkus, W. Samek, S. Lapuschkin, M. M. C.</p>
      <p>Höhne, Quantus: An explainable ai toolkit for responsible evaluation of neural network
explanations and beyond, 2023. URL: https://arxiv.org/abs/2202.06861. arXiv:2202.06861.
[24] L. Arras, A. Osman, W. Samek, Clevr-xai: A benchmark dataset for the ground truth evaluation of
neural network explanations, Information Fusion 81 (2022) 14–40. URL: http://dx.doi.org/10.1016/
j.infus.2021.11.008. doi: 10.1016/j.inffus.2021.11.008.
[25] J. Gildenblat, contributors, Pytorch library for cam methods, https://github.com/jacobgil/
pytorch-grad-cam, 2021.
[26] H. Rezatofighi, N. Tsoi, J. Gwak, A. Sadeghian, I. Reid, S. Savarese, Generalized intersection over
union: A metric and a loss for bounding box regression, 2019. URL: https://arxiv.org/abs/1902.09630.
arXiv:1902.09630.
[27] E. Sizikova, N. Saharkhiz, D. Sharma, M. Lago, B. Sahiner, J. G. Delfino, A. Badano,
Knowledgebased in silico models and dataset for the comparative evaluation of mammography ai for a range of
breast characteristics, lesion conspicuities and doses, Advances in Neural Information Processing
Systems (2023).
[28] O. Russakovsky, J. Deng, H. Su, J. Krause, S. Satheesh, S. Ma, Z. Huang, A. Karpathy, A. Khosla,
M. Bernstein, A. C. Berg, L. Fei-Fei, Imagenet large scale visual recognition challenge, 2015. URL:
https://arxiv.org/abs/1409.0575. arXiv:1409.0575.
[29] M. Raghu, T. Unterthiner, S. Kornblith, C. Zhang, A. Dosovitskiy, Do vision transformers see like
convolutional neural networks?, CoRR abs/2108.08810 (2021). URL: https://arxiv.org/abs/2108.
08810. arXiv:2108.08810.</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <ref id="ref1">
        <mixed-citation>
          [1]
          <string-name>
            <given-names>A.</given-names>
            <surname>Dosovitskiy</surname>
          </string-name>
          ,
          <string-name>
            <given-names>L.</given-names>
            <surname>Beyer</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A.</given-names>
            <surname>Kolesnikov</surname>
          </string-name>
          ,
          <string-name>
            <given-names>D.</given-names>
            <surname>Weissenborn</surname>
          </string-name>
          ,
          <string-name>
            <given-names>X.</given-names>
            <surname>Zhai</surname>
          </string-name>
          ,
          <string-name>
            <given-names>T.</given-names>
            <surname>Unterthiner</surname>
          </string-name>
          ,
          <string-name>
            <given-names>M.</given-names>
            <surname>Dehghani</surname>
          </string-name>
          ,
          <string-name>
            <given-names>M.</given-names>
            <surname>Minderer</surname>
          </string-name>
          , G. Heigold,
          <string-name>
            <given-names>S.</given-names>
            <surname>Gelly</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J.</given-names>
            <surname>Uszkoreit</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Houlsby</surname>
          </string-name>
          ,
          <article-title>An image is worth 16x16 words: Transformers for image recognition at scale</article-title>
          ,
          <year>2021</year>
          . URL: https://arxiv.org/abs/
          <year>2010</year>
          .11929. arXiv:
          <year>2010</year>
          .11929.
        </mixed-citation>
      </ref>
      <ref id="ref2">
        <mixed-citation>
          [2]
          <string-name>
            <given-names>A.</given-names>
            <surname>Vaswani</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Shazeer</surname>
          </string-name>
          ,
          <string-name>
            <given-names>N.</given-names>
            <surname>Parmar</surname>
          </string-name>
          ,
          <string-name>
            <given-names>J.</given-names>
            <surname>Uszkoreit</surname>
          </string-name>
          ,
          <string-name>
            <given-names>L.</given-names>
            <surname>Jones</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A. N.</given-names>
            <surname>Gomez</surname>
          </string-name>
          , Ł. Kaiser,
          <string-name>
            <surname>I. Polosukhin</surname>
          </string-name>
          ,
          <article-title>Attention is all you need</article-title>
          ,
          <source>Advances in neural information processing systems</source>
          <volume>30</volume>
          (
          <year>2017</year>
          ).
        </mixed-citation>
      </ref>
      <ref id="ref3">
        <mixed-citation>
          [3]
          <string-name>
            <given-names>R.</given-names>
            <surname>Azad</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A.</given-names>
            <surname>Kazerouni</surname>
          </string-name>
          ,
          <string-name>
            <given-names>M.</given-names>
            <surname>Heidari</surname>
          </string-name>
          ,
          <string-name>
            <given-names>E. K.</given-names>
            <surname>Aghdam</surname>
          </string-name>
          ,
          <string-name>
            <given-names>A.</given-names>
            <surname>Molaei</surname>
          </string-name>
          ,
          <string-name>
            <given-names>Y.</given-names>
            <surname>Jia</surname>
          </string-name>
          ,
          <string-name>
            <surname>A</surname>
          </string-name>
          . Jose,
          <string-name>
            <given-names>R.</given-names>
            <surname>Roy</surname>
          </string-name>
          ,
          <string-name>
            <given-names>D.</given-names>
            <surname>Merhof</surname>
          </string-name>
          ,
          <article-title>Advances in medical image analysis with vision transformers: A comprehensive review</article-title>
          ,
          <source>Medical Image Analysis</source>
          <volume>91</volume>
          (
          <year>2024</year>
          )
          <article-title>103000</article-title>
          . URL: https://www.sciencedirect.com/science/article/pii/ S1361841523002608. doi:https://doi.org/10.1016/j.media.
          <year>2023</year>
          .
          <volume>103000</volume>
          .
        </mixed-citation>
      </ref>
    </ref-list>
  </back>
</article>